Original Paper
Abstract
Background: Measurement of heart rate (HR) through an unobtrusive, wrist-worn optical HR monitor (OHRM) could enable earlier recognition of patient deterioration in low acuity settings and enable timely intervention.
Objective: The goal of this study was to assess the agreement between the HR extracted from the OHRM and the gold standard 5-lead electrocardiogram (ECG) connected to a patient monitor during surgery and in the recovery period.
Methods: In patients undergoing surgery requiring anesthesia, the HR reported by the patient monitor’s ECG module was recorded and stored simultaneously with the photopletysmography (PPG) from the OHRM attached to the patient’s wrist. The agreement between the HR reported by the patient’s monitor and the HR extracted from the OHRM’s PPG signal was assessed using Bland-Altman analysis during the surgical and recovery phase.
Results: A total of 271.8 hours of data in 99 patients was recorded simultaneously by the OHRM and patient monitor. The median coverage was 86% (IQR 65%-95%) and did not differ significantly between surgery and recovery (Wilcoxon paired difference test P=.17). Agreement analysis showed the limits of agreement (LoA) of the difference between the OHRM and the ECG HR were within the range of 5 beats per minute (bpm). The mean bias was –0.14 bpm (LoA between –3.08 bpm and 2.79 bpm) and –0.19% (LoA between –5 bpm to 5 bpm) for the PPG- measured HR compared to the ECG-measured HR during surgery; during recovery, it was –0.11 bpm (LoA between –2.79 bpm and 2.59 bpm) and –0.15% (LoA between –3.92% and 3.64%).
Conclusions: This study shows that an OHRM equipped with a PPG sensor can measure HR within the ECG reference standard of –5 bpm to 5 bpm or –10% to 10% in the perioperative setting when the PPG signal is of sufficient quality. This implies that an OHRM can be considered clinically acceptable for HR monitoring in low acuity hospitalized patients.
doi:10.2196/27765
Keywords
Introduction
Timely recognition of deterioration in hospitalized patients is important because early intervention improves clinical outcomes of mortality and unplanned intensive care unit (ICU) admissions and reduces length of stay [
]. Especially in perioperative care, complications related to surgery limit effectiveness of the surgery and are associated with increased mortality and costs [ , ]. From previous studies, it is known that vital signs such as heart rate (HR) and respiratory rate are important indicators of critical illness and are often altered long before a deterioration is clinically apparent [ - ]. In general, patients’ vital signs are assessed multiple times a day in general wards. However, patients may deteriorate between the scheduled measurements [ ]. Therefore, both remote and continuous monitoring of HR and respiratory rate is considered a promising tool for early detection of patient deterioration in the low acuity or home setting.The gold standard for measurement of HR in the perioperative setting is the multiple-lead electrocardiogram (ECG). However, there are practical limitations to continuous measurements of vital signs using ECG due to the obtrusiveness and limited mobility of patients. Novel solutions to monitor vital signs have been proposed in the literature [
]. One of these novel solutions is the wrist-based optical heart rate monitor (OHRM). The OHRM has the advantage of offering unobtrusive, remote, and continuous monitoring. The photopletysmography (PPG) sensor in the OHRM has shown potential to provide robust peak detection from which HR may be calculated [ , ]. Validation studies have been presented on the accuracy of these devices in healthy participants [ - ]. However, it remains unclear whether these tools are also reliable for monitoring vital signs in patients during hospital stay. The robustness of an OHRM should be studied in hospitalized patients before it can be reliably adopted in a clinical setting. Few studies have been performed in hospitalized patients, and these included mainly stable ward patients [ , , ]. To check the accuracy of the OHRM in the acute phases of disease, the study population should ideally experience some deterioration in HR during the study period. Hospitalized patients are a heterogeneous population where HR can be influenced by all kinds of pathologies, particularly during surgery, which induces hemodynamic, metabolic, endocrine, and immunological alterations [ , ]. The objective of this study was to assess the agreement between the HR extracted from a PPG sensor–based OHRM and that of the gold standard 5-lead ECG connected to the patient monitor during surgery and recovery.Methods
Study Design
We used a prospective, nonrandomized, observational, single-center study design to examine the perioperative period. The study was performed in the Catharina Hospital in Eindhoven, the Netherlands, a tertiary hospital that performs an average of 20,000 surgical procedures annually. The study was reviewed and approved by the Medical Research Ethics Committees United (study #NL65134.100.18).
Study Population
All adult patients scheduled for noncardiac surgery were screened by anesthesiologists for inclusion in the study. Patients were selected by the anesthesiologist on a weekly basis and informed of the study prior to the surgical procedure. In total, 203 patients were eligible for inclusion, and 100 patients signed informed consent. Cardiac surgeries were excluded since the required extracorporeal circulation and scheduled ICU admission would complicate analysis.
To obtain a representative case mix of patients undergoing surgery, patients were categorized and stratified based on the American Society of Anesthesiologists Physical Status Classification (ASA class) [
] and risk of the surgery [ ]. Patients were divided into 2 groups: (1) low risk (ASA score I or II and low- or intermediate-risk surgery) and (2) high risk (ASA score III or IV and intermediate- or high-risk surgery). If the ASA score and risk were discordant (eg, ASA score IV and low-risk surgery), the ASA score took precedence over the surgical risk.Study Procedure
The measurements on the OHRM started as soon as the device was placed on the patient’s wrist in the holding area. The choice of wrist depended on the placement of the blood pressure cuff. Unless not otherwise possible, the OHRM was placed on the wrist of the arm opposite to the blood pressure cuff to prevent disturbance in the optical measurements of the cardiac pulse. The vital sign measurement started upon arrival in the operating room when sensor modules were connected to the patient monitor. Measurements continued during surgery (surgical phase). After completion of the surgery, the patient was disconnected from the patient monitor located in the operating room and transferred to the recovery room. Upon arrival in the recovery room, the patient monitor was reconnected to the patient monitor located in the recovery room, and measurements continued (recovery phase) until the patient was transferred to the general ward. Upon transfer, the patient monitor was disconnected, and the OHRM was removed from the patient’s wrist.
Data Collection
The wrist-worn OHRM was developed by Philips and equipped with a Philips Cardio and Motion Monitoring Module, which integrates a PPG and accelerometer sensor (
). PPG is an optical technique used to detect volumetric changes in blood in peripheral circulation. It continuously measures the reflectivity of the skin in the green part of the light spectrum in combination with the 3-axial acceleration of the body part where it is located. Accelerometry is a technique used to quantify movement patterns through the detection of rotational and translational acceleration. The sampling frequency of both the PPG and accelerometer sensors was 32 Hz [ ]. The patient monitor in both the operating and recovery room was a Carescape B850 (GE Healthcare) connected to a 5-lead ECG, pulse oximeter, body temperature sensor, and oscillometric cuff for noninvasive blood pressure measurements or an arterial line for invasive blood pressure measurements. All patient monitors were linked to a patient data collection system which logged data for every patient. The application used for logging data was AnStat (CarePoint). AnStat logs trends and waveforms with a sampling frequency of 100 Hz and events like administration of drugs.Data on patient demographics were extracted from the electronic medical records.
Data Processing
The HR from the 5-lead ECG was derived by the Carescape B850 patient monitor software. The HR from the OHRM was extracted from the logged PPG signal using an algorithm that was previously validated in healthy volunteers in various conditions of rest and physical activity [
]. In brief, the algorithm processed the PPG and motion signal simultaneously to derive HR and a quality index (QI) for the HR measurements with a 1-second interval. Both HR and QI were assessed in real time. The algorithm provided an output every second, but the data were processed using a sliding window of 5 seconds. The HR measurements from the ECG and PPG were synchronized using a cross-correlation function and visual inspection of the resulting overlapped time series. The QI characterized the confidence in the provided metric estimated by the algorithm itself. It was represented on a 5-point scale (from 0-4), where 0 denoted “lowest confidence/output unavailable” and 4 denoted “highest confidence.” The QI was determined by proprietary methods and used to provide a monotonically increasing relation between availability and reliability. The QI of the HR was typically influenced by the signal-to-noise ratio of the PPG signal, the ability of the algorithm to cope with motion artifacts, and the periodicity of the detected pulse signal.A PPG-based arrhythmia detection algorithm [
] was also used to identify periods in which the PPG signal was not in accordance with a normal sinus rhythm. In brief, the arrhythmia detection algorithm first identifies interpulse intervals (IPIs) in a 30-second interval from the PPG signal and then rejects the IPI in presence of motion during the IPI period. The final set of IPIs in the 30-second period are then processed by a Markov Model to define the probability of atrial fibrillation (AF). In our study, if >50% of the detected IPIs in the 30-second interval were rejected by the algorithm, the interval was labeled as arrhythmia. For measurement intervals during which events of arrhythmias were detected by this algorithm, the QI was set to 0. To summarize the PPG signal coverage, each HR measurement was assigned to 1 of 3 categories: (1) good quality (QI=4), (2) low quality (QI≤3), and (3) arrhythmia. Only HR data associated with QI=4 were used in the agreement analysis. Coverage was measured as the ratio between the measurements with good quality and the entire measurement duration for a patient. If patients had less than 5 minutes of coverage during surgery or recovery, the session was excluded from analysis. The hospital health records were screened to find potential causes for patients that were excluded since this would indicate that the OHRM was not usable for these patients.Bland-Altman plots were made to visualize the agreement between ECG and PPG HR [
]. Limits of agreement (LoA) and CIs of the LoA were calculated by taking into account both within- and between-patient variability [ ]. The modified Bland-Altman method that estimates the limits of agreement with repeated measurements where the true value varies, as described by Zou [ ], was used. The CIs of the LoAs were constructed using the method of variance estimate recovery (MOVER). In short, a 1-way random-effects model was used to model the difference dij of the j-th measurement for the i-th patient as follows:dij = d + ai + eij
Where d is the unknown true difference between the ECG and PPG HR. The difference d is either the difference between the PPG and ECG HR (ie, d = HRPPG – HRECG) or the percentage difference calculated by d = d% = . ai and eij are zero-mean normally distributed, with variance and corresponding to the true between- and within-patient variances, respectively. The bias is estimated by , where and and mi is the number of pairs per patient. The between- and within-patient variances were estimated by and where . and were summed to obtain an estimate of the total variance . The 95% LoA values were then calculated by . CIs around the LoA values were estimated by the MOVER [
]. Bland-Altman analysis was conducted for both absolute difference and the percentage difference in HR between PPG and ECG. The HR evaluation was compared to the reference standard [ ], which requires an accuracy of –5 bpm to 5 bpm or –10% to 10% (whichever is largest).Results
Characteristics and Coverage
A total of 100 patients were included. One patient was excluded because the patient monitor data were missing due to technical difficulties. Recovery data of 1 patient were missing because this patient was transferred to the ICU immediately after surgery. Three patients had too few (<5 minutes) good quality PPG measurements during both the surgery and recovery phase and were therefore omitted from the agreement analysis. Another 12 patients had <5 minutes of good quality measurements during either the surgery or recovery phase, but only the respective phase was omitted from the agreement analysis. Patient demographics are shown in
.An example of the data that were captured for each patient in the study is shown in
. In total, 159.08 hours of data were captured during surgery, 76.5% (121.7/159.1 hours) of which were of good quality (QI = 4), and 112.59 hours of data were captured during recovery, 74.4% (83.8/112.6 hours) of which were of good quality. Coverage varied between patients ( ). Median coverage was 86% (IQR 65% to 95%) and did not differ significantly between surgery and recovery (Wilcoxon paired difference test P=.17). Coverage statistics are shown in .Demographic variable | Value | |
Total participants, n | 99 | |
Age in years, median (IQR) | 58.0 (44.5-68.0) | |
Male, n | 36 | |
BMI (kg/m2), median (IQR) | 28.7 (24.8-37.1) | |
ASA-PSa score, n | ||
I | 10 | |
II | 39 | |
III | 45 | |
IV | 5 | |
Surgical risk, n | ||
High | 9 | |
Intermediate | 63 | |
Low | 27 | |
Diabetes, n | 7 | |
Hypertension, n | 37 | |
Hypercholesterolemia, n | 21 | |
Previous stroke or TIAb, n | 13 | |
Structural heart disease, n | 8 | |
Atrial fibrillation, n | 8 | |
Wrist device location, n | ||
Left | 45 | |
Right | 53 | |
Unknown | 1 | |
Surgery type, n | ||
Bariatric | 22 | |
Gastroenterological | 8 | |
Neurological | 3 | |
Orthopedic | 31 | |
Plastic | 7 | |
Thyroid | 1 | |
Urogenital | 17 | |
Vascular | 10 | |
Surgery duration (min), median (IQR) | 87.0 (48.0-115.0) | |
Recovery duration (min), median (IQR) | 58.0 (41.2-78.0) |
aASA-PS: American Society of Anesthesiologists Physical Status.
bTIA: transient ischemic attack.
Characteristics of the collected data | Surgery | Recovery | Surgery and recovery |
Total hours, n | 159.6 | 112.2 | 271.8 |
Good quality PPGa (hours), n (%) | 124.1 (78.0) | 83.8 (74.4) | 207.9 (76.5) |
Low quality PPG (hours), n (%) | 33.3 (21.0) | 28.7 (25.5) | 62.0 (22.8) |
Arrhythmia (hours), n (%) | 1.7 (1.1) | 0.2 (0.2) | 1.9 (0.7) |
aPPG: photopletysmography.
Bland-Altman Analysis During Surgery
The mean bias was –0.15 (SD 0.05) bpm and –0.20% (SD 0.06) for the PPG-measured HR compared to the ECG-measured HR, where the LoA (including the SE) fell within the reference standard of –5 bpm to 5 bpm and –10% to 10% (
).Results | Difference in bpma | Difference in percentage |
Bias, mean (SE) | –0.15 (0.05) | –0.20 (0.06) |
SD of differences | 1.50 | 2.34 |
Lower LoAb (95% CI) | –3.08 (–2.99 to –3.19) | –4.79 (–4.92 to –4.66) |
Upper LoA (95% CI) | 2.79 (2.69 to 2.89) | 4.39 (4.26 to 4.53) |
Within-patient variance | 2.04 | 5.12 |
Between-patient variance | 0.20 | 0.37 |
Intraclass correlation coefficient | 0.09 | 0.07 |
abpm: beats per minute.
bLoA: limits of agreement.
Bland-Altman Analysis During Recovery
The mean bias was –0.10 bpm (SD 0.04) and –0.14% (SD 0.04) for the PPG-measured HR compared to the ECG-measured HR, where the limits of agreement (including the SE) fell within the reference standard of –5 bpm to 5 bpm and –10% to 10% (
).Results | Difference in bpma | Difference in percentage |
Bias, mean (SE) | –0.10 (0.04) | –0.14 (0.04) |
SD of differences | 1.38 | 1.93 |
Lower LoAb (95% CI) | –2.80 (–2.72 to –2.87) | –3.92 (–3.83 to –4.01) |
Upper LoA (95% CI) | 2.59 (2.52 to 2.67) | 3.64 (3.56 to 3.74) |
Within-patient variance | 1.78 | 3.56 |
Between-patient variance | 0.11 | 0.16 |
Intraclass correlation coefficient | 0.06 | 0.04 |
abpm: beats per minute.
bLoA: limits of agreement.
Discussion
A wrist-worn OHRM may be able to provide continuous unobtrusive HR monitoring in the low acuity care or home settings. To determine this, the validity of OHRM-derived HR must first be assessed in a representative target population and compared to the gold standard 5-lead ECG. In this study, the agreement between the HR derived from an OHRM and a 5-lead ECG connected to a patient monitor was assessed for a representative patient population during the perioperative period. The OHRM could provide an accurate HR (–5 bpm to 5 bpm and –10% to 10% compared to the ECG-derived HR) during both the surgical and recovery phase when the PPG signal was of good quality. A vast majority (121.7/159.1 hours, 76.5%) of the PPG signal was good quality.
Given the hemodynamic changes during the perioperative period and the diversity in surgical procedures, a technical validation, as performed in this study, is essential before the OHRM can be introduced into clinical practice. Very few studies were found in the literature that validated wrist-worn OHRMs in hospitalized patients. One study with a goal of early warning detection using an OHRM was performed in patients during and after discharge from the ICU [
]. The OHRM was a personal fitness tracker, and 24 hours of monitoring started in the ICU while patients were still being monitored by means of a continuous ECG. The authors concluded that personal fitness tracker–derived HRs were slightly lower than those derived from continuous ECG monitoring and not as accurate as pulse oximetry-derived HRs. A feasibility study was performed by the same research group regarding bradycardia and tachycardia detection in the same population [ ]. The authors stressed in both studies the importance of subgroup analysis of patients not in sinus rhythm since this negatively impacted measurement accuracy. This corresponds to the findings in our study where measurements during arrhythmia were of low quality.Another study was designed for AF detection, but also showed good results in sinus rhythm in patients undergoing elective cardioversion for AF [
]. There were fewer patients (N=20) included than in our study, and the agreement analysis was based on QRS intervals as the reference, with a mean difference of 1.3 ms being found between ECG and PPG. Other studies were performed in healthy participants and focused on assessing accuracy during physical activity [ , , , , , - ]. However, the results obtained in these studies cannot be translated to our results since surgery was the underlying cause for changes in HR in our study and not physical activity. Factors influencing HR during surgery are hemodynamic changes induced by anesthesia, intraoperative factors such as blood loss and hypothermia, or involvement of vital organs in the area of surgery. Results of previous studies did conclude that motion artifacts remain a challenge in OHRMs. In this study, motion artifacts were less likely to occur since patients were mostly immobilized. Nevertheless, motion artifacts are relevant to consider if the OHRM is to be used in the future for remote monitoring of patients.The agreement between the ECG- and PPG-derived HR was within the LoA of –5 bpm to 5 bpm and –10% to 10% (whichever was largest) both during surgery and recovery. However, this only applied when the quality of the PPG signal was labeled as “good.” Moreover, a vast majority (during surgery: 121.7/159.1 hours; during recovery: 76.5%; 83.8/112.6, 74.4%) of the PPG signal was good quality. Ideally, the coverage should be 100%, but this may not be realistic since a poor signal-to-noise ratio in the PPG measurements can perturb the detection of a sinus rhythm. Arrhythmias such as ectopic beats, AF, premature ventricular or atrial complex, and paced beats also contributed to a reduction of measurement coverage of the OHRM. This is confirmed by the fact that patients with a medical history of AF had lower overall coverage compared to patients without previous diagnosis of AF, resulting in 25% versus 85% overall coverage, respectively. This was also true for those patients with severe congenital heart disease where median coverage was 47% versus 85% for patients without structural heart disease. Finally, a very small group of patients had an extremely low coverage, but a consequently large influence on the mean coverage. Median coverage was higher, with 85% being good quality data. Exclusions of patients in this study should be taken into account as well when clinical applicability of the OHRM is assessed. Furthermore, 3 patients were excluded since <5 minutes of data were captured in total, which could be explained in 1 case by serious congenital heart disease which involved aberrant anatomy. Another 12 patients with <5 minutes of good quality data during surgery or recovery were also excluded. As the reference standard, ECG is considered capable of providing 100% coverage. However, in clinical practice, this is most likely not the case since ECG HR detection can also fail in the presence of the aforementioned abnormalities.
The limitations of this study are the following. Despite a heterogeneous group of elective procedures and hospital setting, no general ward patients were included. Nevertheless, translation of our findings to patients in the general ward is reasonable as patients are transitioning from immobile to a more moveable state during stay in the recovery room. By using a 1-way random-effects model, the between- and within-patient variance was quantified to explore the effect of heterogeneity of the study group. As indicated by Hamilton and Lewis [
], not accounting for repeated measures can lead to a falsely narrow LoA, mainly with a small number of patients and a large number of measurements per patient. Both the mean bias and between-patient variance are weighted according to the number of observations available for each patient. Hence, patients with more observations will contribute more to the final results. As the distribution of observation times was not normal, some patients contributed substantially more than did others, and results could have been biased to these patients. It is also worth mentioning the assumptions underlying the 1-way random-effects model. Specifically, the model assumes that repeated differences on a single patient are independent and that the within-patient variance of these differences is constant and the same for all patients. First, the independence assumption could have been too strong since hemodynamic changes occurred during surgery or recovery which could have led to autocorrelation in the HR and subsequent differences arising between the PPG- and ECG-derived HR. The effect of autocorrelation on the within-patient variance is unknown, and further studies are needed to take autocorrelation into account [ ]. Second, the assumption of homoscedasticity was not formally tested, and it could have been the case that the variance of the differences increased with higher HR. Finally, the possible influence of surgery-specific factors, such as electrosurgical instruments causing interference was not investigated.With this study in hospitalized patients, we gained knowledge on the influence of the oscillometric blood pressure cuff which disturbs the measurements by compromising the blood flow. Although both nurses and patients experience the wireless and unobtrusive wristband as an advantage, the OHRM will still have to find a way into local workflow. Before early warning systems can be incorporated into timely detect deterioration, clinical studies to define the predictive value of continuous HR monitoring for deterioration in hospitalized patients other than the ICU or operating room are first needed. Although our own and previous studies might not have found equal accuracy compared to the gold standard, there is still more opportunity to produce benefit during the acute phases of illness, which otherwise may go unnoticed in the general ward with monitoring of the HR 2 to 3 times daily. Although the use of an ORHM for deterioration detection seems less time-consuming for nurses, it still remains uncertain what the false alarm rate will be or how much time practical procedures, such as charging the battery of the OHRM, will take. From a practical point of view, placement of the wristband is made problematic by intravenous or arterial lines, identification bands, or bandages on the wrists.
In summary, the current study found that the OHRM is clinically acceptable when good quality data are captured and in settings when high-intensity monitoring, such as in the ICU or operating room, is not mandatory. The OHRM seems less suitable for patients with congenital anatomical changes of the heart or patients with arrhythmias. When the OHRM captures a significant amount of low-quality data in a patient, the suggestion would be to use another monitoring type to ensure safe monitoring. Since the OHRM can report the quality of the PPG signal instantaneously, the decision to switch to ECG monitoring can be made immediately. The reliability of an OHRM to measure HR in patients known to suffer from arrhythmias or structural heart disease requires further research.
In conclusion, this study shows that an OHRM equipped with a PPG sensor can measure HR within the ECG reference standard of –5 bpm to 5 bpm and –10% to 10% in the perioperative setting when the PPG signal is of sufficient quality. This implies that an OHRM can be considered clinically acceptable for HR monitoring in low acuity hospitalized patients and may provide the basis for future studies for remote, unobtrusive, continuous monitoring for timely recognition of deterioration.
Acknowledgments
We would like to express our gratitude for the valuable help of Marleen van Aartrijk in the design and collection of data for this study. This research was funded by Rijksdienst voor Ondernemend Nederland, Ewatch (grant # ITEA161006).
Authors' Contributions
RD, RH, JG, VS, HHMK, and RAB conceptualized the study. RD, RS, and RAB devised the methodology. RD and AGB performed the validation. RD conducted the formal analysis. RD and EM were responsible for execution of the study protocol. JG, RD, and EM obtained the resources. AGB and JM were responsible for data curation. RD and EM prepared the original draft. AGB, JM, JG, RH, HHMK, VS, and RAB performed the draft review and editing. RD was responsible for visualization. RD and JG were responsible for project administration. RH and VS were responsible for funding acquisition. All authors have read and agreed to the published version of the manuscript.
Conflicts of Interest
AGB, JM, JG, and RH are employed at Royal Philips Netherlands. RAB and HHMK are clinical consultants for Philips Research Netherlands.
References
- Cardona-Morrell M, Prgomet M, Turner RM, Nicholson M, Hillman K. Effectiveness of continuous or intermittent vital signs monitoring in preventing adverse events on general wards: a systematic review and meta-analysis. Int J Clin Pract 2016 Oct;70(10):806-824. [CrossRef] [Medline]
- Ahmad T, Bouwman RA, Grigoras I, Aldecoa C, Hofer C, Hoeft A, International Surgical Outcomes Study (ISOS) group. Use of failure-to-rescue to identify international variation in postoperative care in low-, middle- and high-income countries: a 7-day cohort study of elective surgery. Br J Anaesth 2017 Aug 01;119(2):258-266 [FREE Full text] [CrossRef] [Medline]
- Noordzij PG, Poldermans D, Schouten O, Bax JJ, Schreiner FAG, Boersma E. Postoperative mortality in The Netherlands: a population-based analysis of surgery-specific risk in adults. Anesthesiology 2010 May;112(5):1105-1115 [FREE Full text] [CrossRef] [Medline]
- Walston JM, Cabrera D, Bellew SD, Olive MN, Lohse CM, Bellolio MF. Vital signs predict rapid-response team activation within twelve hours of emergency department admission. West J Emerg Med 2016 May;17(3):324-330 [FREE Full text] [CrossRef] [Medline]
- Fieselmann JF, Hendryx MS, Helms CM, Wakefield DS. Respiratory rate predicts cardiopulmonary arrest for internal medicine inpatients. J Gen Intern Med 1993 Jul;8(7):354-360. [Medline]
- Taenzer AH, Pyke JB, McGrath SP, Blike GT. Impact of pulse oximetry surveillance on rescue events and intensive care unit transfers: a before-and-after concurrence study. Anesthesiology 2010 Feb;112(2):282-287. [CrossRef] [Medline]
- Suaste-Gómez E, Hernández-Rivera D, Sánchez-Sánchez AS, Villarreal-Calva E. Electrically insulated sensing of respiratory rate and heartbeat using optical fibers. Sensors (Basel) 2014 Nov 14;14(11):21523-21534 [FREE Full text] [CrossRef] [Medline]
- Vadrevu S, Manikandan MS. A robust pulse onset and peak detection method for automated PPG signal analysis system. IEEE Trans. Instrum. Meas 2019 Mar;68(3):807-817. [CrossRef]
- Chang KM. Pulse rate derivation and its correlation with heart rate. Journal of Medical and Biological Engineering 2009;29(3):132-137.
- Sartor F, Papini G, Cox LGE, Cleland J. Methodological shortcomings of wrist-worn heart rate monitors validations. J Med Internet Res 2018 Jul 02;20(7):e10108 [FREE Full text] [CrossRef] [Medline]
- Valenti G, Westerterp K. Optical heart rate monitoring module validation study. : Jan 11-14, 2013; 2013 Presented at: IEEE International Conference on Consumer Electronics (ICCE); 2013; Las Vegas, NV, USA. [CrossRef]
- Gillinov S, Etiwy M, Wang R, Blackburn G, Phelan D, Gillinov AM, et al. Variable accuracy of wearable heart rate monitors during aerobic exercise. Med Sci Sports Exerc 2017 Aug;49(8):1697-1703. [CrossRef] [Medline]
- Kroll RR, Boyd JG, Maslove DM. Accuracy of a wrist-worn wearable device for monitoring heart rates in hospital inpatients: a prospective observational study. J Med Internet Res 2016 Sep 20;18(9):e253 [FREE Full text] [CrossRef] [Medline]
- Delgado-Gonzalo R, Parak J, Tarniceriu A, Renevey P, Bertschi M, Korhonen I. Evaluation of accuracy and reliability of PulseOn optical heart rate monitoring device. 2015 Presented at: The 37th Annual International Conference of the IEEE Engineering in Medicine and Biology Society (EMBC); Aug 25-29, 2015; Milan, Italy. [CrossRef]
- Parak J, Korhonen I. Evaluation of wearable consumer heart rate monitors based on photopletysmography. 2014 Presented at: The 36th Annual International Conference of the IEEE Engineering in Medicine and Biology Society; Aug; Aug, 26-30, 2014; Chicago, IL, USA. [CrossRef]
- Shcherbina A, Mattsson CM, Waggott D, Salisbury H, Christle JW, Hastie T, et al. Accuracy in wrist-worn, sensor-based measurements of heart rate and energy expenditure in a diverse cohort. J Pers Med 2017 May 24;7(2):3-3 [FREE Full text] [CrossRef] [Medline]
- Spierer DK, Rosen Z, Litman LL, Fujii K. Validation of photoplethysmography as a method to detect heart rate during rest and exercise. J Med Eng Technol 2015;39(5):264-271. [CrossRef] [Medline]
- Kroll RR, McKenzie ED, Boyd JG, Sheth P, Howes D, Wood M, et al. Use of wearable devices for post-discharge monitoring of ICU patients: a feasibility study. j intensive care 2017 Nov 21;5(1):64-64. [CrossRef]
- Hochstadt A, Havakuk O, Chorin E, Schwartz AL, Merdler I, Laufer M, et al. Continuous heart rhythm monitoring using mobile photoplethysmography in ambulatory patients. J Electrocardiol 2020;60:138-141. [CrossRef] [Medline]
- Lord JM, Midwinter MJ, Chen Y, Belli A, Brohi K, Kovacs EJ, et al. The systemic immune response to trauma: an overview of pathophysiology and treatment. Lancet 2014 Oct 18;384(9952):1455-1465 [FREE Full text] [CrossRef] [Medline]
- Manou-Stathopoulou V, Korbonits M, Ackland GL. Redefining the perioperative stress response: a narrative review. Br J Anaesth 2019 Nov;123(5):570-583 [FREE Full text] [CrossRef] [Medline]
- Doyle D, Goyal A, Bansal P, Garmon E. American Society of Anesthesiologists Classification (ASA Class). Treasure Island, FL, USA: StatPearls Publishing; 2020.
- Kristensen SD, Knuuti J, Saraste A, Anker S, Bøtker HE, De Hert S, Authors/Task Force Members. 2014 ESC/ESA Guidelines on non-cardiac surgery: cardiovascular assessment and management: The Joint Task Force on non-cardiac surgery: cardiovascular assessment and management of the European Society of Cardiology (ESC) and the European Society of Anaesthesiology (ESA). Eur J Anaesthesiol 2014 Oct;31(10):517-573. [CrossRef] [Medline]
- Papini GB, Fonseca P, van Gilst MM, Bergmans JWM, Vullings R, Overeem S. Wearable monitoring of sleep-disordered breathing: estimation of the apnea–hypopnea index using wrist-worn reflective photoplethysmography. Sci Rep 2020 Aug 11;10(1):13512. [CrossRef]
- Kathirvel P, Sabarimalai Manikandan M, Prasanna SRM, Soman KP. An efficient R-peak detection based on new nonlinear transformation and first-order Gaussian differentiator. Cardiovasc Eng Tech 2011 Oct 12;2(4):408-425. [CrossRef]
- Bonomi AG, Schipper F, Eerikäinen LM, Margarito J, van Dinther R, Muesch G, et al. Atrial fibrillation detection using a novel cardiac ambulatory monitor based on photo-plethysmography at the wrist. J Am Heart Assoc 2018 Aug 07;7(15):e009351 [FREE Full text] [CrossRef] [Medline]
- Bland JM, Altman D. Statistical methods for assessing agreement between two methods of clinical measurement. Lancet 1986 Feb 08;1(8476):307-310. [Medline]
- Bland JM, Altman DG. Agreement between methods of measurement with multiple observations per individual. J Biopharm Stat 2007;17(4):571-582 [FREE Full text] [CrossRef] [Medline]
- Zou GY. Confidence interval estimation for the Bland-Altman limits of agreement with multiple observations per individual. Stat Methods Med Res 2013 Dec;22(6):630-642 [FREE Full text] [CrossRef] [Medline]
- ANSI/AAMI. Cardiac monitors, heart rate meters,alarms, american national standard (Revision of ANSI/AAMI EC13). Association for the Advancement of Med. Instrum, Arlington, VA. Tech. Rep 2002:1.
- Hochstadt A, Chorin E, Viskin S, Schwartz AL, Lubman N, Rosso R. Continuous heart rate monitoring for automatic detection of atrial fibrillation with novel bio-sensing technology. J Electrocardiol 2019;52:23-27. [CrossRef] [Medline]
- Dobbs WC, Fedewa MV, MacDonald HV, Holmes CJ, Cicone ZS, Plews DJ, et al. The accuracy of acquiring heart rate variability from portable devices: a systematic review and meta-analysis. Sports Med 2019 Mar;49(3):417-435. [CrossRef] [Medline]
- Wallen MP, Gomersall SR, Keating SE, Wisløff U, Coombes JS. Accuracy of heart rate watches: implications for weight management. PLoS One 2016 May;11(5):e0154420 [FREE Full text] [CrossRef] [Medline]
- Wang R, Blackburn G, Desai M, Phelan D, Gillinov L, Houghtaling P, et al. Accuracy of wrist-worn heart rate monitors. JAMA Cardiol 2017 Jan 01;2(1):104-106. [CrossRef] [Medline]
- Hamilton C, Lewis S. The importance of using the correct bounds on the Bland-Altman limits of agreement when multiple measurements are recorded per patient. J Clin Monit Comput 2010 Jun;24(3):173-175. [CrossRef] [Medline]
Abbreviations
AF: atrial fibrillation |
ASA-PS: American Society of Anesthesiologists Physical Status |
bpm: beats per minute |
ECG: electrocardiography |
HR: heart rate |
ICU: intensive care unit |
IPI: interpulse interval |
LoA: limits of agreement |
MOVER: method of variance estimate recovery |
OHRM: optical heart rate monitor |
PPG: photoplethysmography |
QI: quality index |
Edited by G Eysenbach; submitted 05.02.21; peer-reviewed by F Sartor, M O'Brien; comments to author 14.04.21; revised version received 08.06.21; accepted 27.07.21; published 04.11.21
Copyright©Eveline Mestrom, Ruben Deneer, Alberto G Bonomi, Jenny Margarito, Jos Gelissen, Reinder Haakma, Hendrikus H M Korsten, Volkher Scharnhorst, R Arthur Bouwman. Originally published in JMIR Cardio (https://cardio.jmir.org), 04.11.2021.
This is an open-access article distributed under the terms of the Creative Commons Attribution License (https://creativecommons.org/licenses/by/4.0/), which permits unrestricted use, distribution, and reproduction in any medium, provided the original work, first published in JMIR Cardio, is properly cited. The complete bibliographic information, a link to the original publication on https://cardio.jmir.org, as well as this copyright and license information must be included.