Continuous mHealth Patch Monitoring for the Algorithm-Based Detection of Atrial Fibrillation: Feasibility and Diagnostic Accuracy Study

Continuous mHealth Patch Monitoring for the Algorithm-Based Detection of Atrial Fibrillation: Feasibility and Diagnostic Accuracy Study

Continuous mHealth Patch Monitoring for the Algorithm-Based Detection of Atrial Fibrillation: Feasibility and Diagnostic Accuracy Study

Original Paper

1School of Medicine, Faculty of Health Sciences, University of Eastern Finland, Kuopio, Finland

2Doctoral School, Faculty of Health Sciences, University of Eastern Finland, Kuopio, Finland

3Department of Applied Physics, University of Eastern Finland, Kuopio, Finland

4Centre for Prehospital Emergency Care, Kuopio University Hospital, Kuopio, Finland

5Heart Center, North Karelia Central Hospital, Joensuu, Finland

6Department of Clinical Physiology and Nuclear Medicine, Kuopio University Hospital, Kuopio, Finland

7Department of Emergency Medicine and Services, Helsinki University Hospital and University of Helsinki, Helsinki, Finland

8Heart Center, Kuopio University Hospital, Kuopio, Finland

9Department of Emergency Care, Kuopio University Hospital, Kuopio, Finland

*these authors contributed equally

Corresponding Author:

Onni E Santala, MD

School of Medicine

Faculty of Health Sciences

University of Eastern Finland

Yliopistonranta 1 PO BOX 1627

Kuopio, FI-70211

Finland

Phone: 358 503010879

Email: elmeris@uef.fi


Background: The detection of atrial fibrillation (AF) is a major clinical challenge as AF is often paroxysmal and asymptomatic. Novel mobile health (mHealth) technologies could provide a cost-effective and reliable solution for AF screening. However, many of these techniques have not been clinically validated.

Objective: The purpose of this study is to evaluate the feasibility and reliability of artificial intelligence (AI) arrhythmia analysis for AF detection with an mHealth patch device designed for personal well-being.

Methods: Patients (N=178) with an AF (n=79, 44%) or sinus rhythm (n=99, 56%) were recruited from the emergency care department. A single-lead, 24-hour, electrocardiogram-based heart rate variability (HRV) measurement was recorded with the mHealth patch device and analyzed with a novel AI arrhythmia analysis software. Simultaneously registered 3-lead electrocardiograms (Holter) served as the gold standard for the final rhythm diagnostics.

Results: Of the HRV data produced by the single-lead mHealth patch, 81.5% (3099/3802 hours) were interpretable, and the subject-based median for interpretable HRV data was 99% (25th percentile=77% and 75th percentile=100%). The AI arrhythmia detection algorithm detected AF correctly in all patients in the AF group and suggested the presence of AF in 5 patients in the control group, resulting in a subject-based AF detection accuracy of 97.2%, a sensitivity of 100%, and a specificity of 94.9%. The time-based AF detection accuracy, sensitivity, and specificity of the AI arrhythmia detection algorithm were 98.7%, 99.6%, and 98.0%, respectively.

Conclusions: The 24-hour HRV monitoring by the mHealth patch device enabled accurate automatic AF detection. Thus, the wearable mHealth patch device with AI arrhythmia analysis is a novel method for AF screening.

Trial Registration: ClinicalTrials.gov NCT03507335; https://clinicaltrials.gov/ct2/show/NCT03507335

JMIR Cardio 2022;6(1):e31230

doi:10.2196/31230

Keywords



Background

Atrial fibrillation (AF) is globally the most common arrhythmia associated with significant morbidity and mortality, imposing a significant burden on patients, public health, and the health care budget [1,2]. The detection of AF is important in the decision to initiate anticoagulation therapy to prevent thromboembolic events [3,4]. Nonetheless, AF detection is still a major clinical challenge as AF is often paroxysmal and asymptomatic [5-8]. AF screening recommendations include opportunistic or systematic screening in patients ≥65 years of age or in those individuals with other characteristics pointing to an increased risk of stroke [3]. The European Heart Rhythm Association recommends multiple time points or over a prolonged time to increase diagnostic yield in AF screening with digital devices [9]. The popularities of well-being and taking personal responsibility for one’s own health are reflected in the continuous development and growth of mobile health (mHealth) technologies. There are currently more than 400 wearable activity monitors and more than 100,000 mHealth apps already available [10]. mHealth technologies could provide an additional opportunity to diagnose AF, particularly its paroxysmal and asymptomatic forms [11]. Most mHealth technologies designed for AF detection include some form of automatic AF detection algorithm [12]. These novel mHealth technologies could provide a cost-effective solution for AF-screening [13].

Objective

A highly popular and evolving mHealth technology could reach the population to be screened at a relatively low cost and with little logistical effort, since well-being and personal health care monitoring is already a part of everyday life in many individuals. Several of these new devices produce heart rate variability (HRV) data, which have been widely used to assess a variety of well-being measures such as recovery and stress monitoring [14]. Furthermore, HRV monitoring could also be suitable for other health-related measurements such as AF detection [10,15]. Algorithm-based rhythm monitoring with devices designed for well-being and health-related measurements could enable straightforward, cost-effective, and reliable methods for AF screening. The specific aims of this study were to (1) evaluate the feasibility and quality of the HRV data using a single-lead electrocardiogram (ECG)–based mHealth patch device, and (2) assess the accuracy of an artificial intelligence (AI) arrhythmia detection algorithm in AF screening with 24-hour monitoring.


Study Setting and Participant Recruitment

The study was conducted as a single-center study between April 2018 and December 2019 in Kuopio University Hospital. The study patients were recruited from the hospital emergency care department. The inclusion criteria were AF or sinus rhythm (SR) based on a 12-lead resting ECG recorded during admission to the hospital. The exclusion criteria were as follows: (1) estimated stay in the hospital <24 hours; (2) BMI ≥35kg/m2; (3) left bundle branch block or right bundle branch block; (4) implanted cardiac pacemaker; and (5) a medical condition requiring immediate treatment. The clinical characteristics and symptoms prior to hospital admission of the patients were collected using a standardized data collection protocol and confirmed or complemented from the medical records. All participants provided written informed consent to participate in the study. All data were pseudonymized. Each subject was given an ID number, and the data were kept locked and encrypted in the study files. The measurement data did not include personal data. The research data collected in the study will be treated confidentially as required by the Personal Data Act.

HRV Measurements and Analysis

A single-lead ECG-based HRV device (Firstbeat Bodyguard 2, Firstbeat Technologies) was applied to the patient’s chest with 2 adhesive patches as shown in Figure 1. The Firstbeat Bodyguard 2 device records ECG data, from which it stores beat-to-beat R-R intervals (time elapsed between two successive R-waves) to allow an HRV assessment. The target time for HRV measurement was 24 hours. Simultaneously registered 3-lead Holter ECG recording (Faros 360, Bittium) was used as the “gold standard” for rhythm classification (Figure 1). Both devices stored the data in the internal memory of the device.

Figure 1. Heart rate variability (HRV) and electrocardiogram (ECG) recordings. Single-lead ECG-based HRV recording (1) and 3-lead Holter ECG recording (2). LA: left arm; LL: left limb; RA: right arm; V3: V3 lead in 12-lead ECG.
View this figure

A commercial arrhythmia analysis software (Awario, Heart2Save) was used for automatic AF screening from the HRV data. The AI arrhythmia detection algorithm classified the HRV data in 30-second time windows into 3 categories: SR, AF, and uninterpretable. The accuracy of the AI-based rhythm classification from the HRV recording was further assessed by comparing it with the gold standard Holter ECG recording. Holter ECG recordings were analyzed using a Medilog Darwin Professional V2.8.1 software (Schiller Global). ECG recordings were reviewed independently by 4 investigators blinded to the initial 12-lead ECG and classified into either AF or non-AF rhythms.

Statistical Analysis

The size of the study sample was estimated as 200 observations with an assumed sensitivity of 95% and with a 3% margin error. The AF and control groups were compared using t test for continuous variables and χ2 test or Fisher exact tests for dichotomous variables. All HRV data were analyzed by the AI arrhythmia detection algorithm in 30-second time windows. The performance of the AI arrhythmia detection algorithm in AF detection from HRV recordings was quantified using accuracy, sensitivity, specificity, negative predictive value (NPV), and positive predictive value (PPV). The performance of the AI arrhythmia detection algorithm was tested by (1) detecting AF per patient (subject-based) and (2) total accumulated AF duration across all patients (time-based). All significance tests were 2-tailed, and P≤.05 was considered statistically significant. The data were analyzed using IBM SPSS statistics software, version 27. This study was registered in the ClinicalTrials.gov database (NCT03507335).

Ethics Approval

The study design was approved by Research Ethics Committee of the Northern Savo Hospital District (347/2018).


Study Population

A total of 654 patients were assessed for eligibility. In the initial assessment for the study patients, 454 (69.4%) patients were excluded for the reasons summarized in Figure 2. A total of 200 eligible patients were included in study, of which 100 (50%) were assigned to the AF group and 100 (50%) with SR to the control group. Of the 200 eligible participants, 22 (11%) were further excluded (n=19, 10% were excluded due to missing data and n=3, 1.5% withdrew their consent). In addition, the rhythm of some patients had converted from the time of 12-lead ECG recording prior to study measurements. Consequently, the final rhythm classification made from Holter ECG recording reclassified 11 patients from the AF group into the control group and 1 patient from the control group into the AF group. Thus, the final study population consisted of 178 patients, of whom 79 (44%) were in the AF group and 99 (56%) in the control group.

Compared to the control group, patients with AF were older (77, SD 10 vs 68, SD 15 years; P<.001), presented more often with a history of paroxysmal AF (P<.001), hypertension (P=.02), congestive heart failure (P<.001), and previous heart surgery (P=.02), and were more often receiving anticoagulation (P<.001), beta-blocker (P<.001), and digoxin (P=.005) therapy. Patients with AF also more often reported palpitations (P=.005) and respiratory distress (P<.001), compared to the control group. These groups did not differ statistically with respect to the other recorded parameters (Table 1).

Figure 2. Study flow chart. AF: atrial fibrillation; HRV: heart rate variability.
View this figure
Table 1. Patient demographics.
Characteristics and demographicsControl group (n=99)AFa group (n=79)Significance (2-sided)
Age (years), mean (SD)68 (15)77 (10)<.001
BMI (kg/m2), mean (SD)26 (4)27 (4).16
Sex, n (%).09

Male, n (%)40 (40)42 (53)

Female59 (60)37 (47)
Medical history, n (%)

Earlier AF episode21 (21)63 (80)<.001

Coronary heart disease27 (27)26 (33).41

Diabetes mellitus21 (21)19 (24).65

Hypertension57 (58)59 (75).02

Congestive heart failure12 (12)39 (49)<.001

Previous heart surgery7 (7)15 (19).02
Medication, n (%)

Anticoagulation therapy26 (26)67 (84)<.001

Beta blocker42 (42)56 (71)<.001

Digoxin4 (4)13 (17).005

Antiarrhythmic medication1 (1)1 (1)>.99
Symptoms prior to hospital admission (%)

Decrease in general condition56 (57)45 (57).96

Fatigue51 (52)47 (60).29

Palpitations23 (23)34 (43).005

Respiratory distress26 (26)39 (49)<.001

Chest pain17 (17)18 (23).35

aAF: atrial fibrillation.

Quality of the HRV Data

In the automatic analysis of HRV data (all subjects), 3099 hours out of 3802 hours (81.5%) were estimated as being interpretable. In Holter ECG recordings, 3723 hours were deemed interpretable. Based on the evaluation of the AI arrhythmia detection algorithm, 1308/1543 hours (84.8%) of the AF group recordings and 1753/2180 hours (80.4%) of the control group recordings were estimated as being interpretable. The subject-based median for interpretable data was 99% (25th percentile=77% and 75th percentile=100%) (Figure 3).

Figure 3. Percentage of interpretable heart rate variability data in individual subject recordings. Subjects sorted by using automatic quality values.
View this figure

Accuracy of the AI Arrhythmia Detection Algorithm When Interpreting HRV Data

The accuracy of rhythm analyses was evaluated from the data deemed interpretable in both HRV and ECG recordings (3062 hours). The AI arrhythmia algorithm detected AF correctly in all the patients in the AF group and suggested the presence of AF in 5 patients in the control group (false-positive AF detection), resulting in a subject-based AF detection accuracy of 97.2% (173/178), a sensitivity of 100% (79/79), a specificity of 94.9% (94/99), PPV of 94.0% (79/84), and NPV of 100% (94/94) (Table 2). The values of the time-based accuracy, sensitivity, and specificity of the AI arrhythmia detection algorithm for identifying AF were 98.7% (3022/3062 hours), 99.6% (1304/1308 hours) and 98.0% (1718/1753 hours), respectively; the PPV was 97.4% (1304/1339 hours) for detecting the presence of AF, and the NPV was 99.7% (1718/1723 hours) for its absence (Table 3). We inspected for the reasons for false-positive AF detections, which were as follows: frequent (>10,000 per 24 hours) ventricular extrasystoles, atrioventricular block, a short section (12 minutes) with many ventricular extrasystoles, and in 2 cases they were attributable to undetected QRS complexes (Q wave, R wave, and S wave).

Table 2. Subject- and time-based AF detection accuracy, sensitivity, and specificity based on artificial intelligence arrhythmia detection algorithm.
Algorithm typesAccuracy, n/N (%)Sensitivity, n/N (%)Specificity, n/N (%)
Subject-based algorithm173/178 (97.2)79/79 (100)94/99 (94.9)
Time-based algorithm (h)3022/3062 (98.7)1304/1308 (99.6)1718/1753 (98.0)
Table 3. PPVa and NPVb based on artificial intelligence arrhythmia detection algorithm.
Algorithm typesPPV, n/N (%)NPV, n/N (%)
Subject-based algorithm79/84 (94.0)94/94 (100)
Time-based algorithm (h)1304/1339 (97.4)1718/1723 (99.7)

aPPV: positive predictive value.

bNPV: negative predictive value.


Principal Findings

We demonstrated the good feasibility and diagnostic performance of the mHealth patch device and AI arrhythmia detection algorithm for AF detection. The main findings were as follows: (1) the mHealth patch device designed for well-being and personal health care measurements produced interpretable HRV data from 24-hour recording; and (2) the AI arrhythmia detection algorithm achieved an accurate AF detection from the HRV data.

Comparison to Prior Work

ECG-based wearables, especially adhesive patches, are probably the most valuable approach for tracking HRV data [16]. In previous studies using adhesive patch devices, 92%-99% of the ECG or ECG-based HRV data have been reported to be analyzable with recording times ranging from 12 hours to several days [17-19]. In line with these reports, in our study of the HRV data produced by the ECG-based mHealth patch, 82% were analyzable, and in the subject-based analysis, the median of the data to be analyzed was 99%. However, interindividual differences were observed in the HRV data quality. In 32/178 patients (18%), less than 50% of the data were interpretable (Figure 3). According to a recent systematic review, although the correlation between HRV as measured by Holter and ECG-based wearable devices was excellent at rest, it declined down to 0.85 during movement or exercise [20]. In our study, mobilization of the study patients was not restricted. Despite this, the mHealth device provided high-quality ECG data for HRV monitoring.

The diagnostic accuracy of the AI arrhythmia detection algorithm used in this study to detect AF was comparable to other screening methods and devices. In the previous studies, the sensitivity of automatic AF detection ranged from 67% to 100% and the specificity from 84% to 100% depending on the mobile or digital technology and strategy used [21-36]. In our study, the sensitivity and specificity of AF detection were 100% and 94.9%, respectively. Although the new mHealth screening methods have a high sensitivity and specificity for AF detection, fals-positive AF detections (false alarm) remain a major concern, especially when low-risk individuals use these devices [37]. A false alarm can lead to anxiety, additional health care costs, and possibly even inappropriate treatment [37]. Nonetheless, it should be noted that some false alarms may be due to another arrhythmia or some other pathological condition, and in these cases, screened individuals may benefit from the AF screening program [38]. Our results support this idea since 3 (60%) of the 5 “false alarms” in our study were due to a true arrhythmia. Further, the threshold for confirming the arrhythmia diagnosis by a health care professional from an ECG recording should be low.

Future Directions

The current guidelines recommend AF screening for individuals at an elevated risk of stroke, such as those over 65 years of age [3]. However, there is no unambiguous guidance on the actual screening strategy to be adopted. According to the current guidelines of the European Society of Cardiology, the development of wearable technology is likely to provide cost-effective and practical alternatives for the detection of AF [3]. The popularity of well-being and personal health care monitoring is reflected in the continuous development and growth of the mHealth technologies [10]. However, most of these techniques have not been validated for clinical purposes [10]. Therefore, before using mHealth technologies, one should carefully consider whether the technology in question has been validated for clinical use [14]. Novel methods, such as the AI arrhythmia analysis used here, could provide reliable AF screening when combined with technologies designed for well-being and personal health care measurements. Firstly, a longer duration of rhythm monitoring is more sensitive compared to single recording in AF screening [39]. Secondly, the advantage of automatic AF screening helps to identify arrhythmia episodes from long-term ECG or HRV data, particularly in patients with paroxysmal and asymptomatic AF. Consequently, these techniques could enable the screening of patients at high risk for AF and stroke and the allow follow-up of patients after cardioversion and ablation as well as those recovering from stroke or transient ischemic attack.

Study Limitations

We acknowledge some limitations in our study. First, the presence of morbid obesity could degrade the signal quality and thus produce more failed measurements. In addition, in the right bundle branch block and left bundle branch block cases, the presence of broad 2-peak QRS complexes could result in a false R-R interval, and thereby, an erroneous AF detection by the automatic algorithm. For these reasons, these patients were excluded, and further studies will be needed in these subgroups. In the future, algorithm-based detection of AF with HRV mHealth patch devices will need to be conducted in an out-of-hospital setting to assess the signal quality and accuracy of AF detection in these scenarios.

Conclusions

The mHealth patch device provided good quality HRV data for accurate automatic AF detection from 24-hour monitoring. The AI arrhythmia detection algorithm detected AF with high accuracy, sensitivity, and specificity. Thus, the wearable mHealth patch device with AI arrhythmia analysis could represent a new, promising method for AF screening.

Acknowledgments

This work was supported by the Research Committee of the Kuopio University Hospital Catchment Area for the State Research Funding (project 5101137 and project 507T044). OES received research support from the Helsinki University Hospital fund, Department of Emergency Medicine and Services, Antti and Tyyne Soininen Foundation, Aarne Koskelo Foundation, The Finnish Foundation for Cardiovascular Research, and The Finnish Medical Foundation. We thank the nursing and medical staff of the Kuopio University Hospital and especially Menna Kuosmanen, Matias Rapo, Antti Redsven, Roope Sepponen, Salla Autio, and Vilppu Kekkonen for their work with data acquisition, Oona Silvennoinen for illustration assistance, and Tuomas Selander for statistical advice.

Authors' Contributions

All authors contributed to the study and preparation of this manuscript. TJM, JAL, HJ, MPT, JH, JEKH, TTR, OES, and ESV designed the study. OES and ESV were responsible for collecting the data. JAL, OES, ESV, TPL, and TML contributed to data analysis and figure and table preparation. OES, JH, HJ, TTR, MPT, TPL, TML, MC, ESV, OAR, NSN, JEKH, JAL, and TJM were responsible for editing and further improvements to the manuscript. OES was responsible for the literature search, drafting the first version of the manuscript, final editing, and preparation of the manuscript for submission. All authors read and approved the final manuscript.

Conflicts of Interest

OES, JAL, TTR, TJM, HJ, JH, ESV, and MPT are shareholders of a company (Heart2Save) that designs electrocardiogram-based software for medical equipment. JAL, MPT, and HJ report personal fees from Heart2Save.

  1. Benjamin EJ, Muntner P, Alonso A, Bittencourt MS, Callaway CW, Carson AP, American Heart Association Council on EpidemiologyPrevention Statistics CommitteeStroke Statistics Subcommittee. Heart disease and stroke statistics-2019 update: a report from the American Heart Association. Circulation 2019 Mar 05;139(10):e56-e528 [FREE Full text] [CrossRef] [Medline]
  2. Chugh SS, Havmoeller R, Narayanan K, Singh D, Rienstra M, Benjamin EJ, et al. Worldwide epidemiology of atrial fibrillation: a Global Burden of Disease 2010 Study. Circulation 2014 Feb 25;129(8):837-847 [FREE Full text] [CrossRef] [Medline]
  3. Hindricks G, Potpara T, Dagres N, Arbelo E, Bax JJ, Blomström-Lundqvist C, ESC Scientific Document Group. 2020 ESC Guidelines for the diagnosis and management of atrial fibrillation developed in collaboration with the European Association for Cardio-Thoracic Surgery (EACTS): The Task Force for the diagnosis and management of atrial fibrillation of the European Society of Cardiology (ESC) Developed with the special contribution of the European Heart Rhythm Association (EHRA) of the ESC. Eur Heart J 2021 Feb 01;42(5):373-498. [CrossRef] [Medline]
  4. January CT, Wann LS, Calkins H, Chen LY, Cigarroa JE, Cleveland JC, et al. 2019 AHA/ACC/HRS Focused Update of the 2014 AHA/ACC/HRS Guideline for the Management of Patients With Atrial Fibrillation: A report of the American College of Cardiology/American Heart Association Task Force on Clinical Practice Guidelines and the Heart Rhythm Society in collaboration with the Society of Thoracic Surgeons. Circulation 2019 Jul 09;140(2):e125-e151 [FREE Full text] [CrossRef] [Medline]
  5. Potpara TS, Polovina MM, Marinkovic JM, Lip GYH. A comparison of clinical characteristics and long-term prognosis in asymptomatic and symptomatic patients with first-diagnosed atrial fibrillation: the Belgrade Atrial Fibrillation Study. Int J Cardiol 2013 Oct 12;168(5):4744-4749. [CrossRef] [Medline]
  6. Akao M, Chun Y, Wada H, Esato M, Hashimoto T, Abe M, Fushimi AF Registry Investigators. Current status of clinical background of patients with atrial fibrillation in a community-based survey: the Fushimi AF Registry. J Cardiol 2013 Apr;61(4):260-266 [FREE Full text] [CrossRef] [Medline]
  7. Boriani G, Laroche C, Diemberger I, Fantecchi E, Popescu MI, Rasmussen LH, et al. Asymptomatic atrial fibrillation: clinical correlates, management, and outcomes in the EORP-AF Pilot General Registry. Am J Med 2015 May;128(5):509-18.e2. [CrossRef] [Medline]
  8. Siontis KC, Gersh BJ, Killian JM, Noseworthy PA, McCabe P, Weston SA, et al. Typical, atypical, and asymptomatic presentations of new-onset atrial fibrillation in the community: Characteristics and prognostic implications. Heart Rhythm 2016 Jul;13(7):1418-1424 [FREE Full text] [CrossRef] [Medline]
  9. Svennberg E, Tjong F, Goette A, Akoum N, Di Biase L, Bordachar P, et al. How to use digital devices to detect and manage arrhythmias: an EHRA practical guide. Europace 2022 Apr 03:Online ahead of print. [CrossRef] [Medline]
  10. Li KHC, White FA, Tipoe T, Liu T, Wong MC, Jesuthasan A, et al. The current state of mobile phone apps for monitoring heart rate, heart rate variability, and atrial fibrillation: narrative review. JMIR Mhealth Uhealth 2019 Feb 15;7(2):e11606 [FREE Full text] [CrossRef] [Medline]
  11. Giebel GD, Gissel C. Accuracy of mHealth devices for atrial fibrillation screening: systematic review. JMIR Mhealth Uhealth 2019 Jun 16;7(6):e13641 [FREE Full text] [CrossRef] [Medline]
  12. Zungsontiporn N, Link MS. Newer technologies for detection of atrial fibrillation. BMJ 2018 Oct 17;363:k3946. [CrossRef] [Medline]
  13. Freedman B, Camm J, Calkins H, Healey JS, Rosenqvist M, Wang J, AF-Screen Collaborators. Screening for atrial fibrillation: a report of the AF-SCREEN International Collaboration. Circulation 2017 May 09;135(19):1851-1867. [CrossRef] [Medline]
  14. Peake JM, Kerr G, Sullivan JP. A critical review of consumer wearables, mobile applications, and equipment for providing biofeedback, monitoring stress, and sleep in physically active populations. Front Physiol 2018;9:743 [FREE Full text] [CrossRef] [Medline]
  15. Kelli HM, Witbrodt B, Shah A. The future of mobile health applications and devices in cardiovascular health. Euro Med J Innov 2017 Jan;2017:92-97 [FREE Full text] [Medline]
  16. Dobbs WC, Fedewa MV, MacDonald HV, Holmes CJ, Cicone ZS, Plews DJ, et al. The accuracy of acquiring heart rate variability from portable devices: a systematic review and meta-analysis. Sports Med 2019 Mar 31;49(3):417-435. [CrossRef] [Medline]
  17. Breteler MJM, Huizinga E, van Loon K, Leenen LPH, Dohmen DAJ, Kalkman CJ, et al. Reliability of wireless monitoring using a wearable patch sensor in high-risk surgical patients at a step-down unit in the Netherlands: a clinical validation study. BMJ Open 2018 Feb 27;8(2):e020162 [FREE Full text] [CrossRef] [Medline]
  18. Turakhia MP, Hoang DD, Zimetbaum P, Miller JD, Froelicher VF, Kumar UN, et al. Diagnostic utility of a novel leadless arrhythmia monitoring device. Am J Cardiol 2013 Aug 15;112(4):520-524 [FREE Full text] [CrossRef] [Medline]
  19. Solomon MD, Yang J, Sung SH, Livingston ML, Sarlas G, Lenane JC, et al. Incidence and timing of potentially high-risk arrhythmias detected through long term continuous ambulatory electrocardiographic monitoring. BMC Cardiovasc Disord 2016 Feb 17;16:35 [FREE Full text] [CrossRef] [Medline]
  20. Georgiou K, Larentzakis AV, Khamis NN, Alsuhaibani GI, Alaska YA, Giallafos EJ. Can wearable devices accurately measure heart rate variability? a systematic review. Folia Med (Plovdiv) 2018 Mar 01;60(1):7-20. [CrossRef] [Medline]
  21. Chan P, Wong C, Poh YC, Pun L, Leung WW, Wong Y, et al. Diagnostic performance of a smartphone-based photoplethysmographic application for atrial fibrillation screening in a primary care setting. J Am Heart Assoc 2016 Jul 21;5(7):e003428 [FREE Full text] [CrossRef] [Medline]
  22. Lau JK, Lowres N, Neubeck L, Brieger DB, Sy RW, Galloway CD, et al. iPhone ECG application for community screening to detect silent atrial fibrillation: a novel technology to prevent stroke. Int J Cardiol 2013 Apr 30;165(1):193-194. [CrossRef] [Medline]
  23. Lowres N, Neubeck L, Salkeld G, Krass I, McLachlan AJ, Redfern J, et al. Feasibility and cost-effectiveness of stroke prevention through community screening for atrial fibrillation using iPhone ECG in pharmacies. The SEARCH-AF study. Thromb Haemost 2014 Jun;111(6):1167-1176. [CrossRef] [Medline]
  24. Himmelreich JC, Karregat EP, Lucassen WA, van Weert HC, de Groot JR, Handoko ML, et al. Diagnostic accuracy of a smartphone-operated, single-lead electrocardiography device for detection of rhythm and conduction abnormalities in primary care. Ann Fam Med 2019 Sep 09;17(5):403-411 [FREE Full text] [CrossRef] [Medline]
  25. Brasier N, Raichle CJ, Dörr M, Becke A, Nohturfft V, Weber S, et al. Detection of atrial fibrillation with a smartphone camera: first prospective, international, two-centre, clinical validation study (DETECT AF PRO). Europace 2019 Jan 01;21(1):41-47 [FREE Full text] [CrossRef] [Medline]
  26. Orchard J, Lowres N, Freedman SB, Ladak L, Lee W, Zwar N, et al. Screening for atrial fibrillation during influenza vaccinations by primary care nurses using a smartphone electrocardiograph (iECG): A feasibility study. Eur J Prev Cardiol 2016 Oct;23(2 suppl):13-20. [CrossRef] [Medline]
  27. Cunha S, Antunes E, Antoniou S, Tiago S, Relvas R, Fernandez-Llimós F, et al. Raising awareness and early detection of atrial fibrillation, an experience resorting to mobile technology centred on informed individuals. Res Social Adm Pharm 2020 Jun;16(6):787-792. [CrossRef] [Medline]
  28. Tieleman RG, Plantinga Y, Rinkes D, Bartels GL, Posma JL, Cator R, et al. Validation and clinical use of a novel diagnostic device for screening of atrial fibrillation. Europace 2014 Sep;16(9):1291-1295 [FREE Full text] [CrossRef] [Medline]
  29. Vaes B, Stalpaert S, Tavernier K, Thaels B, Lapeire D, Mullens W, et al. The diagnostic accuracy of the MyDiagnostick to detect atrial fibrillation in primary care. BMC Fam Pract 2014 Jun 09;15:113 [FREE Full text] [CrossRef] [Medline]
  30. Doliwa PS, Frykman V, Rosenqvist M. Short-term ECG for out of hospital detection of silent atrial fibrillation episodes. Scand Cardiovasc J 2009 Jun;43(3):163-168. [CrossRef] [Medline]
  31. Proesmans T, Mortelmans C, Van Haelst R, Verbrugge F, Vandervoort P, Vaes B. Mobile phone-based use of the photoplethysmography technique to detect atrial fibrillation in primary care: diagnostic accuracy study of the FibriCheck app. JMIR Mhealth Uhealth 2019 Mar 27;7(3):e12284 [FREE Full text] [CrossRef] [Medline]
  32. Tison GH, Sanchez JM, Ballinger B, Singh A, Olgin JE, Pletcher MJ, et al. Passive detection of atrial fibrillation using a commercially available smartwatch. JAMA Cardiol 2018 May 01;3(5):409-416 [FREE Full text] [CrossRef] [Medline]
  33. Santala OE, Lipponen JA, Jäntti H, Rissanen TT, Halonen J, Kolk I, et al. Necklace-embedded electrocardiogram for the detection and diagnosis of atrial fibrillation. Clin Cardiol 2021 May;44(5):620-626 [FREE Full text] [CrossRef] [Medline]
  34. Väliaho E, Kuoppa P, Lipponen JA, Martikainen TJ, Jäntti H, Rissanen TT, et al. Wrist band photoplethysmography in detection of individual pulses in atrial fibrillation and algorithm-based detection of atrial fibrillation. Europace 2019 Jul 01;21(7):1031-1038. [CrossRef] [Medline]
  35. Hartikainen S, Lipponen JA, Hiltunen P, Rissanen TT, Kolk I, Tarvainen MP, et al. Effectiveness of the chest strap electrocardiogram to detect atrial fibrillation. Am J Cardiol 2019 May 15;123(10):1643-1648. [CrossRef] [Medline]
  36. Santala OE, Halonen J, Martikainen S, Jäntti H, Rissanen TT, Tarvainen MP, et al. Automatic mobile health arrhythmia monitoring for the detection of atrial fibrillation: prospective feasibility, accuracy, and user experience study. JMIR Mhealth Uhealth 2021 Oct 22;9(10):e29933 [FREE Full text] [CrossRef] [Medline]
  37. Ding EY, Marcus GM, McManus DD. Emerging technologies for identifying atrial fibrillation. Circ Res 2020 Jun 19;127(1):128-142 [FREE Full text] [CrossRef] [Medline]
  38. Welton NJ, McAleenan A, Thom HH, Davies P, Hollingworth W, Higgins JP, et al. Screening strategies for atrial fibrillation: a systematic review and cost-effectiveness analysis. Health Technol Assess 2017 May;21(29):1-236. [CrossRef] [Medline]
  39. Diederichsen SZ, Haugan KJ, Kronborg C, Graff C, Højberg S, Køber L, et al. Comprehensive evaluation of rhythm monitoring strategies in screening for atrial fibrillation: insights from patients at risk monitored long term with an implantable loop recorder. Circulation 2020 May 12;141(19):1510-1522. [CrossRef] [Medline]


AF: atrial fibrillation
AI: artificial intelligence
ECG: electrocardiogram
HRV: heart rate variability
mHealth: mobile health
NPV: negative predictive value
PPV: positive predictive value
SR: sinus rhythm


Edited by T Leung; submitted 14.06.21; peer-reviewed by D Filos, O Pavliuk; comments to author 06.12.21; revised version received 27.12.21; accepted 02.05.22; published 21.06.22

Copyright

©Onni E Santala, Jukka A Lipponen, Helena Jäntti, Tuomas T Rissanen, Mika P Tarvainen, Tomi P Laitinen, Tiina M Laitinen, Maaret Castrén, Eemu-Samuli Väliaho, Olli A Rantula, Noora S Naukkarinen, Juha E K Hartikainen, Jari Halonen, Tero J Martikainen. Originally published in JMIR Cardio (https://cardio.jmir.org), 21.06.2022.

This is an open-access article distributed under the terms of the Creative Commons Attribution License (https://creativecommons.org/licenses/by/4.0/), which permits unrestricted use, distribution, and reproduction in any medium, provided the original work, first published in JMIR Cardio, is properly cited. The complete bibliographic information, a link to the original publication on https://cardio.jmir.org, as well as this copyright and license information must be included.