Abstract
Background: Smart textiles (ie, electronic textiles) offer a promising solution to ease continuous electrocardiogram (ECG) monitoring, but their real-world clinical application has been limited.
Objective: This review comprehensively examines the current state of research on textile-based ECG monitoring systems, synthesizing current evidence with respect to performance (ie, signal quality, function under static and dynamic conditions), user experience, and current challenges.
Methods: A systematic literature search across the PubMed, MEDLINE, and Embase databases from 2000 to 2025 identified 34 research papers eligible for inclusion.
Results: Textile-based ECG electrodes demonstrated good signal quality and comfort, particularly under static conditions. Nonetheless, integration into clinical practice requires addressing critical issues, which include greater efforts at validating these technologies in clinical settings and populations, as well as ensuring data security, cost‑effectiveness, user‑friendliness, and data interoperability.
Conclusions: Considering the prominence of feasibility research, the successful clinical integration of textile-based ECG monitoring systems requires comprehensive efforts at establishing a clinical evaluation research base (via clinical trials) and developing regulatory policies.
doi:10.2196/74261
Keywords
Introduction
Cardiovascular disease is a leading cause of morbidity and mortality globally, necessitating the development of improved diagnostic and management tools. In recent years, driven by advances in wearable sensors, wireless connectivity, and cloud computing, there have been substantial developments in large-scale ambulatory collection, transmission, and monitoring of various physiological and clinical outcomes [-]. As integrated in telemedicine and remote patient monitoring solutions, these advances offer significant improvements to cardiovascular disease care, where continuous monitoring is crucial for early disease detection and timely intervention [,]. Among wearable technologies, electronic textile sensors have emerged as a promising solution for unobtrusive, continuous physiological monitoring in free-living environments [,], including in ambulatory electrocardiogram (ECG) monitoring [,].
ECG monitoring relies on the detection of the heart’s electrical activity via surface electrodes (placed on the skin), and subsequent transmission of the signal for analysis []. Long-term ECG monitoring is vital for the early detection of cardiac arrhythmias, enabling the prompt treatment and prevention of severe complications. Advanced textile–based ECG systems, often termed “smart clothing” or “smart textiles,” integrate sensors and electronics within wearable garments [-]. These systems offer the potential for continuous ECG monitoring using dry electrodes, which are embedded in conductive fabrics or textiles []. This approach avoids the discomfort and limitations of conventional gel (Ag or AgCl) electrodes and conductive gels [,], providing a convenient method for long-term monitoring [,,]. Smart textiles can additionally reduce the need for invasive monitoring modalities (eg, loop and event monitors) and provide a better diagnostic yield (due to the ease of wearing).
Despite enthusiasm, however, adoption of textile-based ECG monitoring in clinical practice remains limited. While extensive research supports the technology’s technical feasibility in measuring ECG [,,], few clinical studies have assessed the validity of this technology in real‑world settings []. Challenges hindering widespread clinical adoption include the lack of rigorous clinical validation research and compliance with stringent regulatory requirements (eg, data privacy and security, data interoperability with electronic health records [EHRs]). It is important to acknowledge that the successful translation of textile‑based ECG monitoring from research prototypes to widely adopted clinical and consumer applications also hinges on the development of user‑friendly technologies and user acceptance. On the technical front, maintaining adequate electrode‑skin contact pressure for reliable signal acquisition and ensuring the preservation of garment integrity and performance through repeated washing cycles are critical considerations. While addressing these aspects is not the primary focus of this review, technical developments represent active areas of research, closely linked with user comfort and the long‑term viability of textile‑based ECG systems. This review aims to map and synthesize the existing scientific literature on textile‑based ECG monitoring systems, as integrated in clothing and garments.
Methods
This review aims to map the scientific literature on textile-based ECG monitoring systems integrated into clothing and garments. Research studies, published in the English language between 2000 and 2025, that investigated the use of textile‑based wearable technologies for ambulatory ECG monitoring in humans were included in this scoping review. Studies were excluded if they focused solely on nontextile wearables. The search time frame (2000-2025) was chosen to capture recent advances in smart textiles, excluding outdated technologies. The language restriction was applied to ensure feasibility and maintain the consistency of data interpretation. The results were reported according to the PRISMA (Preferred Reporting Items for Systematic Reviews and Meta-Analyses) checklist ().
The search strategy used a combination of keywords and controlled vocabulary terms (MeSH and Emtree) across PubMed, MEDLINE, and EMBASE (via OVID) databases. The details of the search are described in . Key terms included “Smart textile,” “e‑textile,” “smart clothing,” “textile electrode,” “wearable technology,” “biosensors,” “cardiac rehabilitation,” “remote monitoring,” “ambulatory monitoring,” “Holter,” “cardiovascular diseases,” “arrhythmia,” “atrial fibrillation,” “heart disease,” and “cardiac disorders.” Filters for human studies and publication year limits were subsequently applied to ensure adherence with the specified inclusion criteria.
The selection process involved 3 stages. First, all identified records were checked for duplicate publications. Second, following the removal of duplicate records, titles and abstracts of the remaining records were screened for relevance to the research question and the eligibility criteria. Third, full‑text papers of potentially relevant studies were reviewed to confirm eligibility. Any discrepancies were resolved through discussion among the review team members. This process ensured a rigorous selection of studies appropriate for inclusion in the scoping review. In total, across 981 records identified across databases, 63 records were selected for full‑text review of eligibility. Following this review, 34 studies were selected for inclusion ().
Given the limited number of published randomized clinical trials focusing on textile-based ECG in real-world clinical settings, this scoping review included feasibility studies, with a focus on ECG signal quality and user comfort as key outcome measures. This methodological choice acknowledges potential limitations in generalizability. It emphasizes the need for further research that incorporates diverse populations and real-world conditions to fully evaluate the clinical utility of this technology. The limited number of studies reporting on clinical outcomes directly reflects the current state of research in this area. The 34 studies were then comprehensively synthesized with respect to 3 key areas of research in smart textile–based ECG monitoring:
- Comparative signal quality: To evaluate the signal quality of smart textile electrodes to traditional gel electrodes in evaluating various ECG indices (ie, QRS complex, P-wave, T-wave, R-peak).
- Performance in static and dynamic conditions: To evaluate the performance of textile electrodes under static (eg, resting) and dynamic (eg, movement) conditions, examining the impact of motion artifacts on signal quality.
- User experience: To evaluate user comfort, which is a critical factor for long-term adoption and use of smart textile–based ECG monitoring platforms, considering factors such as fabric type, electrode placement, and overall wearability.
Finally, this review outlines a roadmap for key clinical research priorities, together with highlighting the challenges that hinder the wider adoption and integration of textile‑based ECG monitoring systems.

Results
Research Landscape
Between 2000 and 2025, smart textile technologies for ECG monitoring have seen steady progress, with research efforts predominantly based on prototype development [-] and observational research studies [-]. Smart textile technologies reviewed here use a range of conductive materials (eg, silver, copper, stainless steel, and metal inks), as incorporated into different garment types (eg, knitted patches, bands, vests, t-shirts, and bras). Among key design challenges for e‑textiles is reliable electrode adhesion, particularly given the contrasting properties of metallic electrodes and the nonmetallic nature of clothing. To address this, various integration techniques are used, including direct knitting or weaving of electrodes into the fabric, embroidery of conductive yarns, adhesive bonding with specialized durable adhesives, or printing with conductive inks requiring encapsulation. Garment design, such as the use of compression garments, and material selection, prioritizing rough electrode surfaces for improved mechanical interlocking, also play crucial roles. Different electrode materials and garment designs influence the contact pressure between the electrode and the skin. Contact pressure has an important role in ensuring the proper placement of the wearable and providing comfort. Excessive pressure can improve signal quality but can also increase discomfort and cause skin damage, while insufficient pressure can compromise the device’s functionality []. Some studies have investigated compression garments to enhance contact pressure, while others have focused on optimizing electrode design to distribute pressure evenly. While these techniques offer promising solutions, the long-term durability of these adhesion methods and their resistance to repeated washings remains a key focus for future research. Wash durability is a critical consideration for textile-based ECG systems. Repeated washing cycles can degrade conductive materials, alter garment shape, and compromise electrode-skin contact. Research is ongoing to identify durable and washable electrode materials, as well as garment construction techniques that minimize performance degradation over time [].
Although recent reviews, such as that by Xu et al [], have explored the landscape of smart textiles in sports and health care, this work distinguishes itself by providing a focused analysis of textile-based systems specifically engineered for ECG monitoring, with a particular emphasis on clinical translation and utility. As evidenced by the analysis in (static conditions) and (dynamic conditions), the selection of electrode type significantly influences the performance and applicability of these systems. This segmentation by static and dynamic conditions is crucial because static conditions provide a controlled environment to assess basic electrode performance, while dynamic conditions, which introduce motion artifacts and other real-world challenges, are essential for evaluating the true utility of these systems. For example, silver-coated yarns are widely used in static settings, but their susceptibility to motion artifacts can be a limiting factor in dynamic environments. Novel materials, such as conductive polymers, hold promise for improved flexibility and signal stability, though further research is needed to validate their long-term performance in both static and dynamic scenarios. This highlights the importance of carefully considering the intended application and monitoring environment when selecting the optimal electrode technology for textile-based ECG monitoring.
| Author | Study type | Sample size | Electrode type | Findings | Condition |
| Mestrovic et al [] (2007) | Observational | 31 | Textile-based piezoresistive transducer (vest) | Readable signals in cardiac inpatients lying | Static |
| Vojtech et al [] (2013) | Prototype or concept | 1 | Silver-coated polyester and cotton nanoparticles | The amplitude of the individual waves of ECG signal obtained using the textile electrodes was smaller than the reference signal | Static |
| Dai et al [] (2016) | Prototype or concept | 6 | Flexible polypyrrole textile electrodes | Heart rate accuracy of each subject was more than 95% | Static |
| Kannaian et al [] (2012) | Prototype or concept | 3 | Embroidered the conductive yarn on polyester fabric | Electrocardiogram signal from the textile electrode was similar to the gelled electrode | Static |
| Pandian et al [] (2008) | Prototype or concept | 25 | Sensors (silicon rubber with pure silver fillings) | RR intervals presented good stability | Static |
| Tsukada et al [] (2019) | Observational | 66 | T-shirts for men and brassieres for women with textile electrodes | P-wave, QRS, and T-wave were comparable between the textile and conventional electrodes | Static |
| Nigusse et al [] (2020) | Prototype or concept | 6 | Silver-printed textile electrodes | The signals collected by silver-printed textile electrodes showed clear R-peaks without missing and false peaks, indicating good waveforms | Static |
| Arquilla et al [] (2020) | Observational | 8 | Textile electrodes (silver thread) were sewn in a chest-mounted configuration | The sewn electrodes produced clean data, allowing the measurement of beat-to-beat variability and average heart rate | Static |
| Fouassier et al [] (2020) | Feasibility or pilot | 30 | Smart t-shirt composed of 13 textiles electrodes made of silver yarns and hydrogel pads that released water vapor | A sinus rhythm was distinguished for all recordings obtained with both smart textile and reference system for all 3 static conditions | Static |
aRR: R-R interval.
| Author | Study type | Sample size | Electrode type | Findings | Condition |
| Hsu et al [] (2019) | Prototype or concept | 3 | Noncontact dry electrodes in an elastic chest vest and attached with Velcros (not embedded) | ECG signal quality obtained by the noncontact electrode was similar to the conventional gel electrode. ECG signal quality was also good while walking. | Dynamic |
| Paradiso et al [] (2005) | Prototype or concept | 1 | Fabric with strain sensors | Metal-based fabric electrodes reliably recorded bioelectric potentials. | Dynamic |
| Perez de Isla et al [] (2011) | Feasibility or pilot | 31 | Noncontact dry electrodes in an elastic chest vest | Good intermethod agreement for common ECG parameters. Nuubo shirt allowed pretest and posttest echocardiography. The system offers continuous, noninvasive, remote monitoring. The initial device had frequency limitations (0.5-100 Hz), affecting ST-segment analysis. | Dynamic |
| Trindade et al [] (2016) | Prototype or concept | 5 | T-shirts comprised of 5 skin-contact textile electrodes | T-shirt prototypes provided adequate performance in standing states. Motion artifact interference, mainly caused by friction between the textile electrodes and the skin, considerably limited the performance of the prototypes in mobile contexts. | Dynamic |
| Pola et al [] (2007) | Prototype or concept | 4 | Silver electrodes | Jogging caused the biggest problem in measurements. | Dynamic |
| Gonzales et al [] (2015) | Prototype or concept | Not reported | T-shirt with 3 dry silver-based electrodes | Good efficacy of dry electrodes even in a high-motion environment. | Dynamic |
| Li et al [] (2020) | Prototype or concept | 1 | Textile electrodes (silver-coated nylon yarns) knitted into a T-shirt | The smart clothing showed good performance for measuring ECG signals. | Dynamic |
| Ghahjaverstan et al [] (2023) | Feasibility pilot | 20 | Dry electrodes made from conductive yarns, knitted within an elastic chest band | HR measurements comparable to the reference system across all tasks, including exercising. | Dynamic |
| Romagnoli et al [] (2014) | Observational | 12 | Smart textile system (GOW) with embedded textile electrodes in shirt, transmitting data to an electronic module | Excellent agreement between GOW and ECG for RR intervals (LoAs around ±3 ms). Good agreement for MeanRR, SDNN, SD2. Poor agreement for RMSSD, HF, LF/HF, HFnu, and SD1 (wide limits of agreement). The GOW system showed overall good accuracy for HR but limited accuracy for many HRV parameters. | Dynamic |
| Coyle et al [] (2010) | Prototype or concept | Not reported | Piezoresistive sensor | ECG signals gathered during rest showed high quality. | Dynamic |
| Di Rienzo et al [,] (2005, 2006) | Prototype or concept; observational (development phases for the same system) | 1; 31 | Textile-based piezoresistive transducer (vest) | Readable signals in cardiac inpatients pedaling. | Dynamic |
| Di Rienzo et al [] (2013) | Observational | 40 | Textile vest with knitted conductive‑fiber electrodes (single‑lead ECG) | Readable ECG 95.9% vs 82.7% (traditional); artifact rate 4.1% vs 17.3%; arrhythmia detection sensitivity 99.7%, specificity 99.9%; PQ/QRS comparable; RR interval error ~1 ms | Dynamic |
| Weder et al [] (2015) | Prototype or concept | 12 | Two Ag- or Ti-coated polyethylene terephthalate electrodes embedded into a chest belt | The embroidered electrodes presented signals comparable with Ag or AgCl gel electrodes | Dynamic |
| Pani et al [] (2016) | Prototype or concept | 10 | Patch of textile electrodes made by treating conventional fabrics with a highly conductive solution of PEDOT:PSS | Textile electrodes showed similar performance, or even better, in wet conditions | Dynamic |
| Alizadeh Meghrazi et al [] (2020) | Prototype or concept | 6 | Silver-plated nylon yarns and carbon-coated nylon yarns knitted in a waistband | ECG signals were reliably obtained from different locations on the waist | Dynamic |
| Fukuma et al [] (2019) | Feasibility or pilot | 100 | Textile electrodes (Hitoe fabric with PEDOT-PSS) embedded in a t-shirt | The system was suitable for lifestyle or sports activities, demonstrating AF-detection performance similar to that of other wearable devices | Dynamic |
| Pagola et al [] (2023) | Feasibility or pilot | 163 | Textile electrodes in a garment | The textile wearable Holter monitoring detected pAF in 35.37% of patients | Dynamic |
| Teferra et al [] (2021) | Feasibility or pilot | 1 | Textile electrodes embedded in a smart ECG vest | Performance was comparable to a traditional 3-lead Holter monitor | Dynamic |
| Machino et al [] (2023) | Randomized controlled trial | 67 | Dry textile electrodes (PEDOT-PSS and nanofiber) embedded in a garment | Garment ECG (2-week) detected AF recurrence in 18% of patients, significantly higher than 24-hour Holter (6%) | Dynamic |
| Amami et al [] (2022) | Feasibility or pilot | 31 | Silver textile electrodes embedded in an undershirt | Wearable undershirt electrode demonstrated abilities comparable to Holter ECG for ECG monitoring and arrhythmia detection | Dynamic |
| Olmos et al [] (2014) | Observational | 31 | nECG SHIRT: Biomedical shirt with integrated textile electrodes (BlendFix) for ECG signal acquisition | Excellent correlation between Nuubo system and conventional tilt table test for commonly assessed ECG parameters during tilt testing | Dynamic |
| Yu et al [] (2017) | Cohort study | 5 | Wearable 12-lead ECG T-shirt with 10 dry textile electrode patches (Shieldex Medtex P180) and active electrodes | Average per-lead signal coverage ranged from 20.9% to 56.3%. After combining data from all leads (temporal fusion), overall coverage improved to up to 81.9%. A 3-stage artifact detection algorithm effectively identified artifacts from 50-Hz noise and motion. | Dynamic |
| Pagola et al [] (2018) | Feasibility or pilot | 146 | Three-lead ECG vest and 1-lead ECG chest band | Both garments were similarly comfortable during the day and night. However, the vest group presented a longer time of compliance and time analyzed than the chest band. The percentage of missed signal was lower in the vest group. The rate of undiagnosed AF detected with textile Holter was 21.9% | Dynamic |
| Steinberg et al [] (2019) | Prototype or concept | 15 | Bra or shirt with integrated sensors recording a single-lead ECG | The wearable ECG sensor\'s signal quality and accuracy were equivalent to Holter monitoring. Signal coverage of R-R intervals showed a very close overlay between the wearable sensor and Holter signals. The wearable sensor presented high wearing comfort and minimal risk of skin irritation. | Dynamic |
| Neri et al [] (2024) | Observational | 30 | Crop top garment with embedded polymer–based electrodes | YouCare System showed 70% “Good,” 12% “Acceptable,” and 18% “Not Readable” ECG signals | Dynamic |
aPrototype or concept: Early engineering or technical validation studies, often with small samples or bench testing. Feasibility or pilot: Preliminary human studies assessing usability, signal quality, or performance in real-world conditions. Clinical studies: Observational, cohort, or randomized controlled designs.
bHR: heart rate.
cRR: R-R interval.
dLoA: limits of agreement.
eSDNN: standard deviation of NN intervals.
fRMSSD: root mean square of successive differences.
gHF: high-frequency component.
hLF/HF: low-frequency/high-frequency ratio.
iHFnu: normalized high-frequency component.
jHRV: heart rate variability.
kAF: atrial fibrillation.
lpAF: paroxysmal atrial fibrillation.
Comparative Signal Quality of Textile-Based ECG Monitoring and Standard Gel Electrodes
The standard ECG signal comprises morphological features (ie, P-wave, QRS, and T-wave), reflecting atrial and ventricular depolarization and repolarization, and temporal features (ie, PR, QRS, QT, and R-R intervals), representing durations between these events []. These features provide key insights into cardiovascular physiology and pathology, making continuous long-term ECG monitoring essential for detecting cardiovascular disorders and abnormalities [].
Traditionally evaluated over the long term using Holter monitors, ambulatory ECG monitoring relies on adhesive gel (Ag or AgCl) electrodes, which provide excellent signal quality [,]. However, long-term use is often limited by skin irritation, gel drying, and electrode detachment due to perspiration [,,]. Smart textile electrodes offer a potential solution to these issues by using conductive fabrics, materials, or metals to acquire ECG signals without the need for gels [].
The demonstration of the accuracy and reliability of smart textile–based ECG monitoring platforms is essential to enhance their adoption in clinical settings. In relation to identifying morphological and temporal ECG features, most studies have focused on the QRS complex, P-wave, T-wave, and R-peak amplitude variations and smart textiles’ signal-to-noise ratio. For example, in an early demonstration of the feasibility of textile-based ECG monitoring, Di Rienzo et al [,] used a conductive fiber vest to accurately detect arrhythmias and heart rate compared to a traditional ECG recorder, while Paradiso et al [] showed comparable results with a metal-yarn vest. Further research investigated various conductive materials (eg, silver-coated nylon, stainless steel yarn []), electrode placement (chest bands, t-shirts, wristbands [,,,]), and different textile constructions, demonstrating good signal quality in many studies in agreement with reference devices [,,,,,,,,-,,,]. However, some studies reported issues, such as noise from dry skin [], signal amplitude discrepancies due to electrode placement [], and waveform irregularities []. While many studies focused on signal quality and accuracy, and some incorporated signal processing algorithms to improve data quality [,,,], few addressed the integration of textile-based ECG data into clinical practice [,-,,-].
Several studies demonstrated good textile-based ECG signal quality [,,,,,,,,,,], highlighting the technology’s promise for long-term monitoring. However, limitations remain: data transmission methods (eg, Bluetooth, Wi-Fi) often require extensive postprocessing; most studies were conducted in controlled laboratory settings with healthy participants and did not use fully integrated wearable form factors; and the clinical validation and long-term feasibility remain insufficiently explored. Future research should address these limitations through more comprehensive real-world studies incorporating diverse populations and integrated clinical workflows to ensure reliable and meaningful data.
Textile-Based ECG Monitoring Performance During Static and Dynamic Conditions
The division of studies based on monitoring setting (static vs dynamic) reveals key trends in the field of textile-based ECG monitoring. As demonstrated in , 9 of the reviewed studies focused on static conditions, likely reflecting the initial stages of development and validation for these technologies. The rationale for this segmentation lies in the fundamentally different challenges presented by each scenario. The static setting allows for the controlled assessment of signal quality and electrode performance, minimizing the confounding effects of motion artifacts and providing a baseline for evaluating the system’s potential. However, the number of studies in the dynamic category demonstrates a growing interest in the application of these systems in real-world scenarios. This shift introduces new challenges, as the dynamic setting often leads to increased motion artifact and greater variability in electrode-skin contact impedance. Examining the types of activities within the dynamic setting (walking, jogging, cycling, etc) further highlights these challenges, with activities involving more vigorous movement generally resulting in lower signal quality.
The inherent flexibility of textile-based ECG systems introduces challenges related to motion artifacts. Di Rienzo et al [] explored the impact of dynamic conditions, examining the impact of motion artifacts on signal quality in telemedicine applications. The electrode-skin interface is particularly susceptible to disruption during dynamic movements, leading to noise and signal degradation, especially with dry or hairy skin [,,,]. Numerous studies have investigated the impact of both static (sitting, lying, standing) and dynamic (eg, walking, jogging, sit-to-stand) activities on ECG signal quality [,,,,,,,,,,,]. For example, Perez de Isla et al [] evaluated a dynamic ECG monitoring system (Nuubo’s dynamic ECG) against a conventional treadmill–based ECG monitoring system during exercise electrography and found comparable performance between the 2 systems on baseline and peak heart rates. Similar results were also seen by Olmos et al [] for heart rate and maximum PR interval on the tilt table test between Nuubo’s technology ECG shirt and conventional ECG monitoring.
In relation to other ECG parameters, Di Rienzo et al [,] also observed adequate QRS complex detection during walking and pedaling, while Pandian et al [] reported good consistency in R-R, QRS, and QT intervals during walking with a textile vest. However, Trindade et al [] noted higher noise amplitude with textile electrodes compared to gel electrodes during walking, and Tsukada et al [] observed increased motion artifacts with trunk twisting. Studies included various strategies to mitigate motion artifacts, including the optimization of electrode placement, use of compression bands to enhance electrode-skin contact, and application of skin moisturizers or humidification techniques [,,]. The use of a wetting device was also shown to improve signal quality in both static and dynamic conditions [].
User Comfort
Wider clinical adoption of textile-based ECG monitoring requires user acceptance, which is linked to comfort and ease of use [,]. While high signal quality and accuracy are crucial, a comfortable and unobtrusive system is essential for encouraging prolonged wear and consistent data acquisition []. However, user comfort has been under-investigated, with only a few studies addressing this critical aspect as a secondary outcome [,,,,].
These studies offer key insights into user perceptions. Pagola et al [] found that patients with stroke preferred the e-textile vest over a conventional chest band for long-term use, with comparable comfort levels across day and night. Wu et al [] reported that an e-textile t-shirt was more comfortable than traditional gel electrodes. Neri et al [], comparing an e-textile–based ECG device to a standard Holter monitor, reported significantly greater patient comfort in the e-textile monitor, compared to the standard monitor. Steinberg et al [] also noted greater comfort and low skin irritation with e-textile sensors, integrated into bras and shirts during 24-hour monitoring, while Arquilla et al [] found no significant comfort difference compared to traditional electrodes. Similarly, Montazeri Ghahjaverstan et al [] reported that all pediatric participants found the textile-based chest band easy to wear, nonirritating, and generally comfortable, with 95% rating the design as appropriate and 75% willing to use it for continuous monitoring at home.
Discussion
This review highlights the significant progress, as well as the challenges faced in optimizing e-textile–based ECG monitoring. Our findings address pertinent issues related to the optimal design and adoption of textile-based ECG monitoring technologies with respect to signal quality, static and dynamic performance, and user comfort.
Signal Quality
In relation to signal quality, the reviewed studies primarily evaluated ECG signal quality of e-textiles in healthy participants within controlled laboratory settings. The current proof-of-concept studies and feasibility testing [,], though essential for establishing minimum viable products, do not address the complexities of adopting e-textile–based ECG monitoring in clinical practice. Therefore, definitive conclusions about the effective use and integration of e-textiles in clinical practice cannot be drawn. Still, these studies achieved reliable ECG signals in e-textiles compared to standard adhesive electrodes and clarified key requirements to achieve reproducible ECG signals in e-textiles. These included consistent electrode placement, stable contact, supported by optimized electrode design, appropriate garment fit, garment care, and wash durability. Overall, these findings point to the need for evaluations of e-textile–based ECG monitoring in everyday life and clinical settings against standard modalities (ie, Holter monitors), incorporating diverse populations, health conditions, and standardized testing protocols.
Static and Dynamic Performance
E-textile–based ECG monitoring is particularly susceptible to motion artifacts, which disrupt ECG signals and degrade signal quality. The majority of studies included monitored ECG in static conditions, likely reflecting the initial validation efforts. While these studies indicated equal-to-superior performance for static e-textile–based ECG assessment [,,,,,], fewer studies included dynamic e-textile–based ECG assessment. Despite the lack of standardized protocols, many studies demonstrated the satisfactory detection of QRS complexes and rhythm abnormalities during moderate dynamic activity [,,,,]. Nonetheless, motion artifacts significantly impacted signal quality during more strenuous activities [,,,]. A limitation of the current literature in relation to dynamic e-textile–based ECG assessment is the lack of standardized protocol, making it difficult to directly compare results across studies. Furthermore, studies in realistic, uncontrolled dynamic environments are still relatively scarce, highlighting the need for further research to address the practical challenges of implementing textile-based ECG monitoring in everyday life. In addition to factors that influence signal quality (mentioned previously), further advancements in signal processing and machine learning are needed to reliably extract clinically meaningful data during dynamic conditions [,,].
User Comfort
Although some feasibility and prototype studies report that textile electrodes may improve comfort compared to traditional gel electrodes [,], evidence remains limited and heterogeneous. Independent studies [,,,] have noted favorable comfort outcomes, while 1 study from our group [] found the textile band to be “generally comfortable.” These findings should be interpreted with caution given the potential for bias and the small sample sizes involved. Overall, comfort remains an underresearched area, and further independent clinical evidence is needed to confirm these preliminary observations. Future research should prioritize standardized assessment of user experience using validated tools and include diverse populations, as comfort and ease of use are critical for long-term adoption.
Clinical Integration
The potential for greater use and clinical integration of e-textiles is promising, yet significant challenges remain. These include interoperability with existing EHR systems, clinical validation, patient privacy, cost-effectiveness, timely feedback mechanisms, and user-friendliness. Effective workflow integration also requires clear standards for data review frequency, protocols for urgent findings, and sustainable reimbursement models. Frameworks, such as the ABCD model []—which evaluates device accuracy, clinical utility, regulatory approval, and cost—can guide adoption. Importantly, these frameworks must incorporate both clinician and patient perspectives [-] to ensure usability and adherence.
Recent clinical investigations provide encouraging evidence of feasibility and diagnostic value in real-world settings. Fukuma et al [] demonstrated that a t-shirt with PEDOT-PSS electrodes was suitable for lifestyle and sports activities, achieving atrial fibrillation (AF) detection performance comparable to other wearables; although this study was included in our review, it primarily involved asymptomatic participants with cardiovascular risk factors rather than patients with established disease. Pagola et al [] reported that intensive 90-day textile Holter monitoring detected paroxysmal AF in 35.37% of patients, underscoring the benefit of extended monitoring. Machino et al [] showed that a 2-week garment ECG detected AF recurrence in 18% of patients—significantly outperforming 24-hour Holter monitoring (6%). Similarly, Amami et al [] found that an undershirt with silver textile electrodes provided ECG monitoring and arrhythmia detection comparable to conventional Holter devices. Collectively, these findings indicate promising feasibility and signal quality; however, conclusions about clinical robustness cannot be drawn without formal quality assessment and larger trials.
Recommendations
The rapid evolution of telemedicine and digital health tools presents significant opportunities to improve cardiovascular disease management [-]. Smart textiles offer a promising pathway toward enhanced remote patient monitoring, providing benefits for patients, health care providers, and the overall health care system. However, successful integration requires addressing several key aspects.
- Wearable data acquisition: This pillar focuses on developing robust and reliable wearable systems capable of accurately monitoring, collecting, and preprocessing ECG data from textile-based sensors. This includes considerations, such as sensor design, electrode placement, and signal quality under various conditions.
- Data transmission and communication: The seamless and secure transmission of ECG data from the wearable device to a remote monitoring center (smartphone or PC) is crucial. This necessitates robust and reliable data communication networks, addressing issues such as signal strength, data security, and bandwidth limitations.
- Cloud-based data analytics and clinical integration: A powerful cloud-based analytics platform is needed to process, analyze, and interpret ECG data efficiently, providing clinicians with readily accessible, accurate, and clinically relevant information [,,]. This platform must facilitate timely alerts for acute events, support offline data access, and seamlessly integrate with existing EHRs. Data processing and analysis techniques are essential components of this system, requiring sophisticated algorithms to filter noise, identify key patterns, and provide meaningful interpretation [,]. The scalability and reliability of the backend infrastructure, including data storage and retrieval, are essential factors determining the system’s performance.
Clinicians require user-friendly clinical portals to visualize and manage this data effectively, enabling timely diagnosis and intervention. Patients and caregivers should have access to relevant information for their self-management and remote health monitoring (). Finally, user training and education are critical, particularly for individuals with low digital literacy []. Addressing these 3 pillars will be key to realizing the full potential of smart textiles in improving cardiovascular care.

Current Issues and Challenges
While textile-based ECG technology shows promise in terms of signal quality, several challenges must be addressed to facilitate its broader adoption in clinical practice. This section outlines key obstacles across 3 interconnected areas: technology, data management, and clinical workflow integration (, ).

| Theme | Challenges for clinical integration | Recommendation |
| Technology |
|
|
| Data |
|
|
| Clinical workflow |
|
|
aECG: electrocardiogram.
bFDA: Food and Drug Administration.
cHIPAA: Health Insurance Portability and Accountability Act.
Technological Challenges
Technological challenges include the following:
- Clinical validation: While numerous studies demonstrate the feasibility of textile-based ECG electrodes in terms of signal quality, none have yet achieved routine clinical adoption (technological readiness level 9). A substantial body of clinical data is needed to validate the technology’s fidelity, interoperability, safety, and effectiveness as intended [].
- Affordability and cost: While smart textiles could potentially reduce health care costs through optimized patient management and early intervention [], this remains unsubstantiated. The cost-effectiveness of this technology needs further investigation, considering both initial investment and long-term operational costs [-]. Reimbursement policies from insurance companies and governments are crucial for broader accessibility [,].
- Usability and acceptability: Limited research exists on user perceptions and acceptance of smart textile ECG monitoring. Factors, such as comfort (thermal, tactile, style, pressure), ease of use, perceived value, and the ability to share data with health care providers, significantly influence user adoption [,,,,]. Further research using user-centered design methodologies (surveys, focus groups, interviews) is essential [,], using frameworks, such as the technology acceptance model to understand user behavior [,].
Data-Related Challenges
Data-related challenges include the following:
- Privacy and security: The collection and transmission of sensitive physiological data through smart textiles raise significant privacy and security concerns []. Robust data protection measures, secure data storage, and transparent data usage policies are essential to safeguard patient information and ensure compliance with relevant regulations [,,]. Informed consent and authorized user access are paramount.
- Interoperability: The seamless integration of smart textile ECG data with existing EHRs and other health care systems is essential for efficient remote patient monitoring [,,]. This requires consistent data formats, standard communication protocols, and a cloud-based architecture that ensures data exchange and interoperability across different service providers [].
- Regulatory policy: The rapid evolution of smart textile technology necessitates a review and update of existing regulatory frameworks []. Clear standards and guidelines are needed to ensure safety and effectiveness, particularly concerning data privacy and security, in accordance with regulations like Health Insurance Portability and Accountability Act and Food and Drug Administration guidelines []. Transparent policies regarding data collection, usage, and sharing are crucial for building trust and facilitating adoption [].
Clinical Workflow Challenges
Clinical workflow challenges include the following:
- Communication and feedback: Effective communication and feedback between patients and health care providers are vital components of remote patient monitoring []. User-friendly clinical dashboards, timely alerts, and efficient reporting mechanisms are required to ensure timely intervention and improve clinical decision-making [,,]. However, challenges include potential delays due to data loss, network issues, and data overload [,].
- Reimbursement: Clear and well‑defined reimbursement pathways are essential to ensure the financial viability of smart‑textile–based ECG monitoring within the health care system [,]. To support adoption, the cost‑effectiveness of these smart wearable systems must be demonstrated, and sustainable reimbursement models need to be developed, taking into account factors such as monitoring duration and clinical service provision [,]. Evidence from clinician surveys also indicates support for a shared‑cost approach, where expenses are distributed among patients, insurance providers, and government payers [].
- Quality improvement: Integrating smart textile data into established clinical workflows enables continuous learning and improvement through feedback loops and data-driven insights [,-]. This requires robust data analysis methods, the development of meaningful clinical indicators, and a strong focus on data-driven decision-making within the health care system. A structured approach to quality improvement, including infrastructural and organizational changes, is crucial for maximizing the benefits of smart textile technology.
Limitations
This scoping review provides a current overview of e-textile–based ECG monitoring but has several important limitations. First, our search strategy did not explicitly include ECG-specific keywords (eg, “ECG,” “EKG,” “Electrocardiogram”), which may have restricted the retrieval of engineering-focused studies emphasizing signal fidelity, such as QRS detection algorithms. While cardiovascular disease terms (eg, “Arrhythmia,” “Atrial Fibrillation”) and monitoring concepts (eg, “Holter,” “ambulatory monitoring”) were expected to capture most relevant literature, omitting explicit ECG terms could have excluded technical performance studies that do not reference disease-specific terminology. Future reviews should incorporate ECG-related keywords and engineering databases to ensure comprehensive coverage of signal processing and hardware design research.
Second, the evidence base is dominated by feasibility and proof-of-concept studies, reflecting the early developmental stage of this technology. These studies often involve small sample sizes and controlled laboratory environments, limiting generalizability to real-world clinical settings. While such controlled conditions are valuable for assessing baseline performance, they may not fully capture usability and reliability under everyday conditions. Furthermore, the absence of standardized testing protocols for key metrics, such as signal quality, wearability, and durability, makes cross-study comparisons challenging. We did not conduct a formal risk of bias assessment, so the findings should be interpreted with caution.
Third, the scope of this review was primarily focused on signal quality, performance under static and dynamic conditions, and user experience. Wash durability, a critical factor for long-term usability, was outside the main scope. Future reviews should evaluate standardized laundering protocols and durability metrics, as these factors are essential for practical implementation and regulatory approval.
Despite these limitations, textile-based ECG monitoring demonstrates compelling potential benefits that conventional ECG methods cannot easily replicate. These include improved access to care for remote or underserved populations, continuous monitoring for the early detection of subtle cardiac changes, enhanced patient comfort and adherence due to increased wearability, and potential cost savings through earlier intervention. While further clinical validation and standardization are necessary, these advantages underscore the transformative potential of smart textiles in cardiac care.
Conclusion
In conclusion, smart textiles offer the potential for the valuable long-term monitoring of ECG parameters, improving the detection of transient events and reducing intervention times. Further, their capacity for unobtrusive, remote monitoring and seamless integration with other health metrics offers unique advantages over conventional ECG. However, widespread support and adoption within clinical workflows depend on addressing the identified limitations, incorporating user feedback, and demonstrating the clinical utility and cost-effectiveness of this technology. Only with reliable and meaningful data can textile-based ECG monitoring serve as a valuable diagnostic tool to guide care, provide a more complete picture of heart health, and improve treatment decisions.
Funding
This study was supported by the MITACS Accelerate Fellowship Program.
Conflicts of Interest
CPS is a former employee of Myant Health Corp. She also held a MITACS internship at Myant Health Corp during her postdoctoral fellowship at the University of Toronto. GC is an employee of Myant Health Corp. BM is a former employee of Myant Health Corp. MP is currently employed by Myant Health Corp. EML is a former employee of Myant Health Corp. MA-M is an employee of Myant Health Corp. SB is a former employee of Myant Health Corp.
References
- Markert C, Sasangohar F, Mortazavi BJ, Fields S. The use of telehealth technology to support health coaching for older adults: literature review. JMIR Hum Factors. Jan 29, 2021;8(1):e23796. [CrossRef] [Medline]
- Malekian Borujeni A, Fathy M, Mozayani N. A hierarchical, scalable architecture for a real-time monitoring system for an electrocardiography, using context-aware computing. J Biomed Inform. Aug 2019;96:103251. [CrossRef] [Medline]
- Soon S, Svavarsdottir H, Downey C, Jayne DG. Wearable devices for remote vital signs monitoring in the outpatient setting: an overview of the field. BMJ Innov. Apr 2020;6(2):55-71. [CrossRef]
- Le K, Narayana H, Servati A, et al. Electronic textiles for electrocardiogram monitoring: a review on the structure–property and performance evaluation from fiber to fabric. Textile Research Journal. Feb 2022;93(3-4):878-910. [CrossRef]
- Nigusse AB, Mengistie DA, Malengier B, Tseghai GB, Langenhove LV. Wearable smart textiles for long-term electrocardiography monitoring—a review. Sensors (Basel). Jun 17, 2021;21(12):4174. [CrossRef] [Medline]
- Sana F, Isselbacher EM, Singh JP, Heist EK, Pathik B, Armoundas AA. Wearable devices for ambulatory cardiac monitoring: JACC state-of-the-art review. J Am Coll Cardiol. Apr 7, 2020;75(13):1582-1592. [CrossRef] [Medline]
- Ashley EA, Raxwal VK, Froelicher VF. The prevalence and prognostic significance of electrocardiographic abnormalities. Curr Probl Cardiol. Jan 2000;25(1):1-72. [CrossRef] [Medline]
- Baig MM, Gholamhosseini H, Connolly MJ. A comprehensive survey of wearable and wireless ECG monitoring systems for older adults. Med Biol Eng Comput. May 2013;51(5):485-495. [CrossRef] [Medline]
- Duncker D, Ding WY, Etheridge S, et al. Smart wearables for cardiac monitoring—real-world use beyond atrial fibrillation. Sensors (Basel). Apr 5, 2021;21(7):2539. [CrossRef] [Medline]
- Stoppa M, Chiolerio A. Wearable electronics and smart textiles: a critical review. Sensors (Basel). Jul 7, 2014;14(7):11957-11992. [CrossRef] [Medline]
- Majumder S, Mondal T, Deen MJ. Wearable sensors for remote health monitoring. Sensors (Basel). Jan 12, 2017;17(1):130. [CrossRef] [Medline]
- Yapici MK, Alkhidir TE. Intelligent medical garments with graphene-functionalized smart-cloth ECG sensors. Sensors (Basel). Apr 16, 2017;17(4):875. [CrossRef] [Medline]
- Alizadeh Meghrazi M, Tian Y, Mahnam A, et al. Multichannel ECG recording from waist using textile sensors. Biomed Eng Online. Jun 16, 2020;19(1):48. [CrossRef] [Medline]
- Teixeira E, Fonseca H, Diniz-Sousa F, et al. Wearable devices for physical activity and healthcare monitoring in elderly people: a critical review. Geriatrics (Basel). Apr 7, 2021;6(2):38. [CrossRef] [Medline]
- Paradiso R, Loriga G, Taccini N. A wearable health care system based on knitted integrated sensors. IEEE Trans Inform Technol Biomed. Sep 2005;9(3):337-344. [CrossRef]
- Di Rienzo M, Rizzo F, Parati G, Brambilla G, Ferratini M, Castiglioni P. MagIC system: a new textile-based wearable device for biological signal monitoring. Applicability in daily life and clinical setting. Conf Proc IEEE Eng Med Biol Soc. 2005;2005:7167-7169. [CrossRef] [Medline]
- Trindade IG, Machado da Silva J, Miguel R, et al. Design and evaluation of novel textile wearable systems for the surveillance of vital signals. Sensors (Basel). Sep 24, 2016;16(10):1573. [CrossRef] [Medline]
- Li M, Xiong W, Li Y. Wearable measurement of ECG signals based on smart clothing. Int J Telemed Appl. 2020;2020:6329360. [CrossRef] [Medline]
- Nigusse AB, Malengier B, Mengistie DA, Tseghai GB, Van Langenhove L. Development of washable silver printed textile electrodes for long-term ECG monitoring. Sensors (Basel). Oct 31, 2020;20(21):6233. [CrossRef] [Medline]
- Pandian PS, Mohanavelu K, Safeer KP, et al. Smart Vest: wearable multi-parameter remote physiological monitoring system. Med Eng Phys. May 2008;30(4):466-477. [CrossRef] [Medline]
- Dai M, Xiao X, Chen X, Lin H, Wu W, Chen S. A low-power and miniaturized electrocardiograph data collection system with smart textile electrodes for monitoring of cardiac function. Australas Phys Eng Sci Med. Dec 2016;39(4):1029-1040. [CrossRef] [Medline]
- Pagola J, Juega J, Francisco-Pascual J, et al. Yield of atrial fibrillation detection with textile wearable Holter from the acute phase of stroke: pilot study of Crypto-AF registry. Int J Cardiol. Jan 15, 2018;251:45-50. [CrossRef] [Medline]
- Steinberg C, Philippon F, Sanchez M, et al. A novel wearable device for continuous ambulatory ECG recording: proof of concept and assessment of signal quality. Biosensors (Basel). Jan 21, 2019;9(1):1-13. [CrossRef] [Medline]
- Hsu CC, Lin BS, He KY, Lin BS. Design of a wearable 12-lead noncontact electrocardiogram monitoring system. Sensors (Basel). Mar 28, 2019;19(7):1509. [CrossRef] [Medline]
- Fouassier D, Roy X, Blanchard A, Hulot JS. Assessment of signal quality measured with a smart 12-lead ECG acquisition T-shirt. Ann Noninvasive Electrocardiol. Jan 2020;25(1):e12682. [CrossRef] [Medline]
- Mestrovic MA, Helmer RJN, Kyratzis L, Kumar D. Preliminary study of dry knitted fabric electrodes for physiological monitoring. Presented at: 2007 3rd International Conference on Intelligent Sensors, Sensor Networks and Information; Dec 3-6, 2007. [CrossRef]
- Pola T, Vanhala J. Textile electrodes in ECG measurement. Presented at: 2007 3rd International Conference on Intelligent Sensors, Sensor Networks and Information; Dec 3-6, 2007. [CrossRef]
- Despang HG, Netz S, Heinig A, Holland HJ, Fischer WJ. Wireless long-term ECG integrated into clothing. Biomed Tech (Berl). Dec 2008;53(6):270-278. [CrossRef] [Medline]
- Kannaian T, Neelaveni R, Thilagavathi G. Design and development of embroidered textile electrodes for continuous measurement of electrocardiogram signals. J Ind Text. Jan 2012;42(3):303-318. [CrossRef]
- Vojtech L, Bortel R, Neruda M, Kozak M. Wearable textile electrodes for ECG measurement. Adv Electr Electron Eng. 2013;11(5):410-414. [CrossRef]
- Gonzales L, Walker K, Keller K, et al. Textile sensor system for electrocardiogram monitoring. 2015. Presented at: 2015 IEEE Virtual Conference on Applications of Commercial Sensors (VCACS); Mar 15 to Oct 15, 2015. [CrossRef]
- Weder M, Hegemann D, Amberg M, et al. Embroidered electrode with silver/titanium coating for long-term ECG monitoring. Sensors (Basel). Jan 15, 2015;15(1):1750-1759. [CrossRef] [Medline]
- Pani D, Dessi A, Saenz-Cogollo JF, Barabino G, Fraboni B, Bonfiglio A. Fully textile, PEDOT: PSS based electrodes for wearable ECG monitoring systems. IEEE Trans Biomed Eng. Mar 2016;63(3):540-549. [CrossRef] [Medline]
- Teferra MN, Hobbs DA, Clark RA, Reynolds KJ. Preliminary analysis of a wireless and wearable electronic-textile EASI-based electrocardiogram. Front Cardiovasc Med. 2021;8:806726. [CrossRef] [Medline]
- Arquilla K, Webb AK, Anderson AP. Textile electrocardiogram (ECG) electrodes for wearable health monitoring. Sensors (Basel). Feb 13, 2020;20(4):1013. [CrossRef] [Medline]
- Tsukada YT, Tokita M, Murata H, et al. Validation of wearable textile electrodes for ECG monitoring. Heart Vessels. Jul 2019;34(7):1203-1211. [CrossRef] [Medline]
- Balsam P, Lodziński P, Tymińska A, et al. Study design and rationale for biomedical shirt-based electrocardiography monitoring in relevant clinical situations: ECG-shirt study. Cardiol J. 2018;25(1):52-59. [CrossRef] [Medline]
- Coyle S, Lau KT, Moyna N, et al. BIOTEX: biosensing textiles for personalised healthcare management. IEEE Trans Inf Technol Biomed. Mar 2010;14(2):364-370. [CrossRef] [Medline]
- Perez de Isla L, Lennie V, Quezada M, et al. New generation dynamic, wireless and remote cardiac monitorization platform: a feasibility study. Int J Cardiol. Nov 17, 2011;153(1):83-85. [CrossRef] [Medline]
- Alcala-Medel J, Michaelson D, Eike RJ, Li Y. Durability study of e-textile electrodes for human body communication. Text Res J. Jan 2024;94(1-2):246-258. [CrossRef]
- Rienzo Di M, Rizzo F, Meriggi P, et al. Applications of a textile-based wearable system for vital signs monitoring. Presented at: Conference Proceedings Annual International Conference of the IEEE Engineering in Medicine and Biology Society; Aug 30 to Sep 3, 2006. [CrossRef]
- Machino T, Aonuma K, Maruo K, et al. Randomized crossover trial of 2-week Garment electrocardiogram with dry textile electrode to reveal instances of post-ablation recurrence of atrial fibrillation underdiagnosed during 24-hour Holter monitoring. PLoS ONE. 2023;18(2):e0281818. [CrossRef] [Medline]
- Montazeri Ghahjaverstan N, Balmer-Minnes D, Taghibeyglou B, et al. Textile-based wearable to monitor heart activity in paediatric population: a pilot study. CJC Pediatr Congenit Heart Dis. Aug 2023;2(4):187-195. [CrossRef] [Medline]
- Amami K, Yoshihisa A, Horikoshi Y, et al. Utility of a novel wearable electrode embedded in an undershirt for electrocardiogram monitoring and detection of arrhythmias. PLoS One. 2022;17(8):e0273541. [CrossRef] [Medline]
- Kim H, Lee S. Evaluation of laundering durability of electro-conductive textile dip-coated on para aramid knit with graphene/waterborne polyurethane composite. Fibers Polym. Nov 2018;19(11):2351-2358. [CrossRef]
- Xu Z, Zhang C, Wang F, et al. Smart textiles for personalized sports and healthcare. Nanomicro Lett. Apr 25, 2025;17(1):232. [CrossRef] [Medline]
- Romagnoli M, Alis R, Guillen J, Basterra J, Villacastin JP, Guillen S. A novel device based on smart textile to control heart’s activity during exercise. Australas Phys Eng Sci Med. Jun 2014;37(2):377-384. [CrossRef] [Medline]
- Di Rienzo M, Racca V, Rizzo F, et al. Evaluation of a textile-based wearable system for the electrocardiogram monitoring in cardiac patients. Europace. Apr 2013;15(4):607-612. [CrossRef] [Medline]
- Fukuma N, Hasumi E, Fujiu K, et al. Feasibility of a T-shirt-type wearable electrocardiography monitor for detection of covert atrial fibrillation in young healthy adults. Sci Rep. Aug 13, 2019;9(1):11768. [CrossRef] [Medline]
- Pagola J, Juega J, Francisco-Pascual J, et al. Intensive 90-day textile wearable Holter monitoring: an alternative to detect paroxysmal atrial fibrillation in selected patients with cryptogenic stroke. Heart Vessels. Jan 2023;38(1):114-121. [CrossRef] [Medline]
- Olmos C, Franco E, Suárez-Barrientos A, et al. Wearable wireless remote monitoring system: an alternative for prolonged electrocardiographic monitoring. Int J Cardiol. Mar 1, 2014;172(1):e43-e44. [CrossRef] [Medline]
- Yu X, Boehm A, Neu W, et al. A wearable 12-lead ECG t-shirt with textile electrodes for unobtrusive long-term monitoring—evaluation of an ongoing clinical trial. 2018. Presented at: EMBEC & NBC 2017; Jun 11-15, 2017:703-706; Tampere, Finland. [CrossRef]
- Neri L, Corazza I, Oberdier MT, et al. Comparison between a single-lead ECG garment device and a Holter monitor: a signal quality assessment. J Med Syst. May 27, 2024;48(1):57. [CrossRef] [Medline]
- Jacob Rodrigues M, Postolache O, Cercas F. Physiological and behavior monitoring systems for smart healthcare environments: a review. Sensors (Basel). Apr 12, 2020;20(8):2186. [CrossRef] [Medline]
- Marozas V, Petrenas A, Daukantas S, Lukosevicius A. A comparison of conductive textile-based and silver/silver chloride gel electrodes in exercise electrocardiogram recordings. J Electrocardiol. 2011;44(2):189-194. [CrossRef] [Medline]
- Di Rienzo M, Meriggi P, Rizzo F, et al. Textile technology for the vital signs monitoring in telemedicine and extreme environments. IEEE Trans Inf Technol Biomed. May 2010;14(3):711-717. [CrossRef] [Medline]
- Di Rienzo M, Rizzo F, Meriggi P, et al. Applications of a textile-based wearable system for vital signs monitoring. Conf Proc IEEE Eng Med Biol Soc. 2006;2006:2223-2226. [CrossRef] [Medline]
- Bayoumy K, Gaber M, Elshafeey A, et al. Smart wearable devices in cardiovascular care: where we are and how to move forward. Nat Rev Cardiol. Aug 2021;18(8):581-599. [CrossRef] [Medline]
- Chatterjee S, Thakur RS, Yadav RN, Gupta L, Raghuvanshi DK. Review of noise removal techniques in ECG signals. IET Signal Process. Dec 2020;14(9):569-590. [CrossRef]
- Smital L, Haider CR, Vitek M, et al. Real-time quality assessment of long-term ECG signals recorded by wearables in free-living conditions. IEEE Trans Biomed Eng. Oct 2020;67(10):2721-2734. [CrossRef] [Medline]
- Serhani MA, T El Kassabi H, Ismail H, Nujum Navaz A. ECG monitoring systems: review, architecture, processes, and key challenges. Sensors (Basel). Mar 24, 2020;20(6):1796. [CrossRef] [Medline]
- Wu W, Pirbhulal S, Sangaiah AK, Mukhopadhyay SC, Li G. Optimization of signal quality over comfortability of textile electrodes for ECG monitoring in fog computing based medical applications. Future Gener Comput Syst. Sep 2018;86:515-526. [CrossRef]
- Nwaru BI, Friedman C, Halamka J, Sheikh A. Can learning health systems help organisations deliver personalised care? BMC Med. Oct 2, 2017;15(1):177. [CrossRef] [Medline]
- Genes N, Violante S, Cetrangol C, Rogers L, Schadt EE, Chan YFY. From smartphone to EHR: a case report on integrating patient-generated health data. NPJ Digit Med. 2018;1:23. [CrossRef] [Medline]
- Shah SS, Gvozdanovic A. Digital health; what do we mean by clinical validation? Expert Rev Med Devices. Dec 2021;18(sup1):5-8. [CrossRef] [Medline]
- Gao Y, Li H, Luo Y. An empirical study of wearable technology acceptance in healthcare. Ind Manag Data Syst. Oct 19, 2015;115(9):1704-1723. [CrossRef]
- Abu-Dalbouh HM. A questionnaire approach based on the technology acceptance model for mobile tracking on patient progress applications. J Comput Sci. Jun 1, 2013;9(6):763-770. [CrossRef]
- Venkatesh V, Thong JYL, Xu X. Consumer acceptance and use of information technology: extending the unified theory of acceptance and use of technology. MIS Q. Mar 1, 2012;36(1):157-178. [CrossRef]
- Molinari G, Molinari M, Di Biase M, Brunetti ND. Telecardiology and its settings of application: an update. J Telemed Telecare. Jun 2018;24(5):373-381. [CrossRef] [Medline]
- Slotwiner DJ, Tarakji KG, Al-Khatib SM, et al. Transparent sharing of digital health data: a call to action. Heart Rhythm. Sep 2019;16(9):e95-e106. [CrossRef] [Medline]
- Yang Z, Zhou Q, Lei L, Zheng K, Xiang W. An IoT-cloud based wearable ECG monitoring system for smart healthcare. J Med Syst. Dec 2016;40(12):286. [CrossRef] [Medline]
- Hill DD, Garcia HM. The integration of patient health record generated from wearable and Internet of Things devices into health information exchanges. Int J Health Med Eng. 2021;15:112-116. URL: https://publications.waset.org/10011846/the-integration-of-patient-health-record-generated-from-wearable-and-internet-of-things-devices-into-health-information-exchanges [Accessed 2026-02-17]
- Kildea J, Battista J, Cabral B, et al. Design and development of a person-centered patient portal using participatory stakeholder co-design. J Med Internet Res. Feb 11, 2019;21(2):e11371. [CrossRef] [Medline]
- Gordon WJ, Landman A, Zhang H, Bates DW. Beyond validation: getting health apps into clinical practice. NPJ Digit Med. 2020;3(1):14. [CrossRef] [Medline]
- Pustiek M, Beristain A, Kos A. Challenges in wearable devices based pervasive wellbeing monitoring. Presented at: 2015 International Conference on Identification, Information, and Knowledge in the Internet of Things (IIKI); Oct 22-23, 2015:236-243; Beijing, China. [CrossRef]
- Chan M, Estève D, Fourniols JY, Escriba C, Campo E. Smart wearable systems: current status and future challenges. Artif Intell Med. Nov 2012;56(3):137-156. [CrossRef] [Medline]
- Karunathilake SP, Ganegoda GU. Secondary prevention of cardiovascular diseases and application of technology for early diagnosis. Biomed Res Int. 2018;2018:5767864. [CrossRef] [Medline]
- Patel S, Park H, Bonato P, Chan L, Rodgers M. A review of wearable sensors and systems with application in rehabilitation. J Neuroeng Rehabil. Apr 20, 2012;9:21. [CrossRef] [Medline]
- Ferguson C, Inglis SC, Breen PP, et al. Clinician perspectives on the design and application of wearable cardiac technologies for older adults: qualitative study. JMIR Aging. Jun 18, 2020;3(1):e17299. [CrossRef] [Medline]
- Bove LA. Increasing patient engagement through the use of wearable technology. J Nurse Pract. Sep 2019;15(8):535-539. [CrossRef]
- Vaportzis E, Clausen MG, Gow AJ. Older adults perceptions of technology and barriers to interacting with tablet computers: a focus group study. Front Psychol. Oct 4, 2017;8:1687. [CrossRef] [Medline]
- Moineau B, Myers M, Ali SS, Popovic MR, Hitzig SL. End-user and clinician perspectives on the viability of wearable functional electrical stimulation garments after stroke and spinal cord injury. Disabil Rehabil Assist Technol. Apr 2021;16(3):241-250. [CrossRef] [Medline]
- Torous JB, Chan SR, Gipson SYMT, et al. A hierarchical framework for evaluation and informed decision making regarding smartphone apps for clinical care. Psychiatr Serv. May 1, 2018;69(5):498-500. [CrossRef] [Medline]
- Kotz D, Gunter CA, Kumar S, Weiner JP. Privacy and security in mobile health: a research agenda. Computer (Long Beach Calif). Jun 2016;49(6):22-30. [CrossRef] [Medline]
- Ndlovu K, Mars M, Scott RE. Interoperability frameworks linking mHealth applications to electronic record systems. BMC Health Serv Res. May 13, 2021;21(1):459. [CrossRef] [Medline]
- Herzeca LF. E-textiles: regulating the future of fashion. Law360; 2016. URL: https://www.gibsondunn.com/wp-content/uploads/documents/publications/Herzeca-E-Textiles-Regulating-The-Future-Of-Fashion-Law360-8-25-16.pdf [Accessed 2026-02-07]
- Accenture Federal Services. Conceptualizing a data infrastructure for the capture, use, and sharing of patient-generated health data in care delivery and research through 2024 draft white paper for a PGHD policy framework. 2016. URL: https://healthit.gov/wp-content/uploads/2026/01/onc_pghd_practical_guide.pdf [Accessed 2026-02-17]
- Bergmo TS. Economic evaluation in telemedicine - still room for improvement. J Telemed Telecare. 2010;16(5):229-231. [CrossRef] [Medline]
- Cortez A, Hsii P, Mitchell E, Riehl V, Smith P. Conceptualizing a data infrastructure for the capture, use, and sharing of patient-generated health data in care delivery and research through 2024. Office of the National Coordinator for Health Information Technology (ONC); 2018. URL: https://healthit.gov/wp-content/uploads/2026/01/onc_pghd_practical_guide.pdf [Accessed 2026-02-07]
- Connecting health and care for the nation: a shared nationwide interoperability roadmap. The Office of the National Coordinator (ONC) for Health Information Technology; 2015. URL: https://ncvhs.hhs.gov/wp-content/uploads/2015/10/Day-2-NCVHS-Dept-Update-POSNACK.pdf [Accessed 2026-02-07]
- Elm JJ, Daeschler M, Bataille L, et al. Feasibility and utility of a clinician dashboard from wearable and mobile application Parkinson’s disease data. NPJ Digit Med. 2019;2(1):95. [CrossRef] [Medline]
- Physician fee schedule. Centers for Medicare & Medicaid Services (CMS). 2021. URL: https://www.cms.gov/medicare/payment/fee-schedules/physician [Accessed 2026-02-07]
- Smuck M, Odonkor CA, Wilt JK, Schmidt N, Swiernik MA. The emerging clinical role of wearables: factors for successful implementation in healthcare. NPJ Digit Med. Mar 10, 2021;4(1):45. [CrossRef] [Medline]
- Manninger M, Zweiker D, Svennberg E, et al. Current perspectives on wearable rhythm recordings for clinical decision-making: the wEHRAbles 2 survey. Europace. Jul 18, 2021;23(7):1106-1113. [CrossRef] [Medline]
- Menear M, Blanchette MA, Demers-Payette O, Roy D. A framework for value-creating learning health systems. Health Res Policy Syst. Aug 9, 2019;17(1):79. [CrossRef] [Medline]
- Mason AR, Barton AJ. The emergence of a learning healthcare system. Clin Nurse Spec. 2013;27(1):7-9. [CrossRef] [Medline]
- Friedman CP, Wong AK, Blumenthal D. Achieving a nationwide learning health system. Sci Transl Med. Nov 10, 2010;2(57):57cm29. [CrossRef] [Medline]
- Friedman CP, Rubin JC, Sullivan KJ. Toward an information infrastructure for global health improvement. Yearb Med Inform. Aug 2017;26(1):16-23. [CrossRef] [Medline]
Abbreviations
| AF: atrial fibrillation |
| ECG: electrocardiogram |
| EHR: electronic health record |
| MeSH : Medical Subject Headings |
| PRISMA: Preferred Reporting Items for Systematic Reviews and Meta-Analyses |
Edited by Andrew Coristine; submitted 20.Mar.2025; peer-reviewed by Amanda Mills, Roman A Surmenev, Sadhasivam Mohanadas; accepted 19.Jan.2026; published 02.Mar.2026.
Copyright© Clarissa Pedrini Schuch, Gabriela Chaves, Bastien Moineau, Sarah Bennett, Meysam Pirbaglou, Edwin Martin Lobo, Milad Alizadeh-Meghrazi. Originally published in JMIR Cardio (https://cardio.jmir.org), 2.Mar.2026.
This is an open-access article distributed under the terms of the Creative Commons Attribution License (https://creativecommons.org/licenses/by/4.0/), which permits unrestricted use, distribution, and reproduction in any medium, provided the original work, first published in JMIR Cardio, is properly cited. The complete bibliographic information, a link to the original publication on https://cardio.jmir.org, as well as this copyright and license information must be included.